51
Installation
of the steam cylinder follows the reverse sequence.
Observe
the following
:
–
Before installing the steam cylinder in the unit, check the Oring of the
cylinder receptacle for damage and replace if necessary.
– Moisten the Oring of the cylinder receptacle with water (do not use
grease or oil), then insert steam cylinder into the receptacle and push
it down to the stop.
– Attach the electrode plugs and the level sensor plug to the respective
electrode and sensor connections according to the following table.
Steam cylinder type RD 422
– Fasten steam hose on the connector of the steam cylinder with hose
clamps.
CAUTION!
– A leaky steam hose can cause damage due to moisture inside the
unit.
– The outlet connector of the steam cylinder is made of plastic.
Do
not overtighten
the hose clamp on the steam connector of the
steam cylinder.
sensor
white
brown
black