![Noraxon TeleMyo 2400T V2 Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/noraxon/telemyo-2400t-v2/telemyo-2400t-v2_user-manual_1700610045.webp)
Noraxon U.S.A., Inc.
TeleMyo 2400T V2
Technical Specifications
Power Requirements:
•
Transmitter – 6 V NiMH rechargeable battery pack. (Noraxon part number: BP2A)
•
Battery charger operates at 100-240 VAC 50/60 Hz
Battery Life:
•
Transmitter – 4 hours (continuous use); Charge time ~ 120 minutes
Output Power & Transmission Frequency (depending on country):
•
Up to 100 mW (depending on antenna and country allowance)
•
DSSS 2412-2462 MHz on (up to) 11 unique radio channels
Transmission Range:
•
Up to 300 feet (100 meters) in open view
•
Antenna options maximize range in specific environments
Dimensions:
•
Transmitter: L: 5.5” x W: 2.3” x H: 1.4” Weight: 14 oz.
•
L: 14cm x W: 5.8cm x H: 3.6cm Weight: 396.9 g.
Hardware Filters:
•
No notch (50/60 Hz) filters are used
•
All sEMG leads have 1
st
order high pass filters set to 10 Hz +/- 10% cutoff
•
All channels have 8
th
order Butterworth/Bessels low pass anti-alias filters set to either 500 Hz or
1000 Hz +/- 2 % cutoff
EMG Active Leads:
•
Baseline noise < 1 uV rms
•
Input Impedance > 100 MegOhm
•
CMR > 100 dB
•
Range +/- 3.5 mV
•
Base gain 500
Transmitter Data Acquisition System:
•
12 bit resolution
•
Input range +/- 5 V (full scale)
•
Noise < 2 LSB
•
8 channels per transmitter
•
All channels have individual gain settings
•
sEMG channel hardware gains (500) x1, x2, x4, x10
•
non sEMG hardware gains x1, x2, x4, x10
•
Sample rates 1500 or 3000 samples/second/channel
P-2308TV2 Rev G (06-06-11)
44