Instruction Manual
NivuFlow 650
page 34
NF 650 - rev. 01 / 29.06.2020
2. Remove the transport protection film from the clear view cover (Fig. 20-1 no. 1) if
available.
Tip:
The protective film will harden upon exposure to UV radiation and possibly cannot
be removed later without leaving some residue. Readability may be strongly impai-
red due to the changes to the protective film.
New clear view covers can be purchased from NIVUS for an extra charge and can
be easily replaced by the user.
3. Install the weatherproof cover if available.
Preparing the field enclosure for electric installation
Â
Ü
Procedure:
1. To remove the clear view cover (Fig. 20-1 no. 1) from
- enclosure type
ZUB0 NFW0
(protection IP67):
Open the four lateral release clamps (Fig. 20-1 no. 2) and remove the enclosure
cover.
- enclosure type
ZUB0 NFW0 IP68 / ZUB0 NFW10 4PFAD
(protection IP68):
Remove the four cylinder head screws M4x25 (Fig. 20-1 no. 7) and the washers,
open the four lateral release clamps (Fig. 20-1 no. 2) and remove the enclosure
cover.
2. To remove the blue inside cover loosen the four round head screws 3.5x25 in the
corners and remove the cover. Now the transmitter including the display (Fig. 20-1
no. 4) and the terminal clamps (Fig. 20-1 no. 5) are fully accessible.
3. The unit is reassembled in reverse order after being wired. Check as well that
- the gaskets are undamaged and free of dirt,
- the screws are firmly tightened.
Otherwise the IP67/IP68 protection degree
cannot be guaranteed
.
Содержание NivuFlow 650
Страница 39: ...Installation and Connection page 39 NF 650 rev 01 29 06 2020 Fig 21 4 Terminal connections NivuFlow 650 Type T4...
Страница 41: ...Installation and Connection page 41 NF 650 rev 01 29 06 2020 Fig 21 6 Terminal connections NivuFlow 650 Type TZ...
Страница 173: ...Approvals and Certificates Page 173 NF 650 rev 01 29 06 2020 Approvals and Certificates...