Thermo Scientific AquaSensors
™
AV88 Universal Analyzer User Guide
30
4.5.
Air Calibration (Sensor Type: D.O. )
An AIR CALIBRATION computes the oxygen content measured in air based on the setting for PRESSURE,
SALINITY and the measured temperature. View or change the Pressure and Salinity settings in the
CONFIGURE menu.
Initiate the AIR CALIBRATION process by pressing ENTER.
Figure 4.11: Air calibrate menu for D.O.
When the air calibration procedure is initiated, the analog output is placed into hold mode. The user is
prompted to place the sensor in air as shown in the Figure 4.12.
Figure 4.12 The user is prompted to place the sensor in air.
After the sensor is placed in air, press the ENTER key to start the calibration.
It takes between 5 seconds and two minutes depending on sensor and temperature measurement stability.
During this time, the calibration procedure can be aborted by pressing the ESC key. See Section 4.2 for an
example of an aborted calibration procedure.
Calibrate
Air
MENU
ESC
ENTER
Air
Sensor in air?
MENU
ESC
ENTER
Air
Calibrating...
MENU
ESC
ENTER