Thermo Scientific AquaSensors
™
AV88 Universal Analyzer User Guide
10
2.3.
Power, Relay and Current Loop Wiring
The AV88 Display comes with two terminal blocks. The 10-pin block on the left is used to hook up two current
loops and two relays (if installed). The 12-pin terminal block on the right is used to hook up 24-volt power, host
communications (if installed) and one of many analytical sensors. There is a sensor personality module for
each sensor type that is plugged in and fastened down as shown if figure 2.1A.
Sensor personality modules show which sensor the AV88 is configured for and provide wiring information.
Figure 2.1A: AV88 wiring when configured for Differential pH
Two Form C
Relays
24 VDC Power
(Digital Comms on pins 3
& 4 when option installed)
Two Current
Loops
Differential pH Sensor
Wiring
Lock nut to retain pH
personality module
Push/Pull rod to
remove pH personality
module
Personality module
label shows pH
parameter, functions
and wire colors.