6.4.5.5
ORP Sensor calibration
Parameter: -
Menu path:
SERVICE / SENSOR CALIBRATION
Description:
Calibration of the ORP sensor.
SET ZERO POINT
Setting the zero point of the sensor
1. Measured value (without setting).
2. Real (measured) value at the zero point of the sensor
Use a 0 mV buffer solution. Dip the sensor into the buffer solution and wait until the measured
value (1) becomes stabilised! Set the uncalibrated measured value (1) in the edit field using
the , and buttons. By pressing and simultaneously the measured value can
be entered to the edit field.
SENSOR SETUP
PARAMETER EDITOR
SET ZERO POINT
3mV mV
0005
3
1
2
Entering the SERVICE / SENSOR CALIBRATION menu the instrument will display a warning message that the user is about to change
critical parameters of the measurement system. The user can enter the menu by pressing the (OK) button or exit to the previous menu by
pressing the (NO) button. In the next dialogue the software asks the user if the calibration of the sensor is the „first” calibration (after a
new sensor is installed) or not. Select OK or NO buttons. If OK button is selected the stored informational parameters of the previous sensor
are reset to default values. Select (NO) button if it is not a „first” setting.
6.4.5.6
DO Sensor calibration
Parameter: -
Menu path:
SERVICE / SENSOR CALIBRATION
Description:
Calibration of the DO sensor. Calibration process is described at chapter 6.6.
Calibration shall be performed in air or referential liquid, sanity is zero in both cases.
CAL. IN AIR (SPAN)
Calibration of saturated value (100%sat) with sensor in air (8.24ppm / [25
˚
C, 760Hgmm])
CAL. IN ZERO SOL.
Calibration of the Zero point in referential liquid.
CAL. IN SPAN SOL.
Calibration of the saturated value (100%sat).
RESET TO DEFAULT
Resets the calibration table to default
RESET TIMER
Resets the inner timer of the pH electrode life-time monitoring routine.
This timer
is NOT a real-time clock
It only counts when the instrument is powered. Its value is saved in a non-volatile memory