Messages
Display
Reason
Temperature
Error
•
Too Cold/
Too Hot
Sample Not
Detected or
Using Wrong
Test Strip
Used Test
Strip
or Test Strip
Outside of
Vial Too Long
Meter
Error
Test Strip Error,
Very high blood
glucose result
(higher than
33.3 mmol/L)
Action
Move Meter and Test
Strips to area between
10°C-40°C; wait 10
minutes for System to
reach room temperature
before testing.
Retest with new
Test Strip and
larger sample.
Repeat with new Test
Strip. Make sure sample
is touched to edge of Test
Strip (not top). If error
persists, call for assistance.
Call for
assistance.
Retest with new Test Strip. If error
persists, call for assistance. If you
have symptoms such as fatigue, excess
urination, thirst, or blurry vision
follow your healthcare professional’s
advice for high blood glucose.