17
How to Test Control Solution
Use ONLY TRUEyou Control Solution
with the TRUEyou Meter.
1. Check supplies. See
Getting Started
.
2. Allow Control Solution, vial of Test
Strips and Meter to adjust to room
temperature (20°C-25°C).
Note:
Running a Control Test at
temperatures outside the range
listed above may cause Control
Solution to read as a blood test.
3. Wash hands. Dry thoroughly.
4. Gently swirl or invert Control Solution bottle to
mix. DO NOT SHAKE!
5. Remove one Test Strip from vial. Close Test
Strip vial immediately. Use Test Strip quickly
after removal from vial.
Note:
If Test Strip has been out of the
vial too long before testing, an
error message appears upon
insertion of the Test Strip into the
Meter. Release and discard old
Test Strip. Use new Test Strip for
testing.
6. Insert Test Strip into Meter. Meter turns on
and shows blinking Drop Symbol and time. Do
not remove Test Strip.