15
Quality Control Testing
To assure you are getting accurate and reliable results,
TRUEyou offers two kinds of quality control tests.
These tests let you know that your TRUEyou System is
working properly and your testing technique is good.
Automatic Self-Test
An Automatic Self-Test ensures that
the Meter is working correctly.
1. Wash hands and dry thoroughly.
2. Remove one Test Strip from the
Test Strip vial and insert Test Strip
into the Meter.
3. Meter turns on. The full Display
appears and is replaced by the time
and the blinking Drop Symbol. The
Meter is working correctly and is ready
to perform a Control or blood test.
Do not use Meter if:
• The full Display does not appear
(segments are missing),
• The blinking Drop Symbol does not appear, or
• An error message appears in the Display.
See
Troubleshooting
or call for assistance (see cover
of this Booklet for phone number).
Full Display
Contacts
Facing Up
Tyou_OB_Guts_E4GBY03FDr20.indd 15
11/8/13 1:43 PM