Trouble shooting
Problem
Possible causes and/or corrective action
B/G LEDs and status LEDs are not
shining
The power cable is not plugged into a power socket or is
not connected to the Power Supply.
B/G LEDs are not shining. Blue LED
is shining
The Transilluminator is in the “heat protection” mode.
Please wait 5 minutes. The LEDs will switch on
automatically.
B/G LEDs are not shining. Blue LED
is blinking
The Transilluminator has switched off. Please use the main
switch to restart the transilluminator. Remember to switch
off the device after work.