![Nikon N-STORM Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/nikon/n-storm/n-storm_operation-manual_1691874016.webp)
Chapter 2
STORM Microscopy Operation
12
3
Finely and properly adjust the laser focal point of the
TIRF illuminator.
(1) Select the checkbox for the 647-nm laser.
(2) Turn on the Perfect Focus System (PFS) in [Ti Pad]
and set the focus.
(3) Finely adjust the laser focal point of the TIRF
illuminator so that the observation target is visible.
4
Check activation.
(1) Make sure that the checkbox for the reporter laser
(647 nm) is selected, and then set the power to
100% by adjusting the power slider.
(2) Make sure that the checkbox for the activation laser
(405 nm, 457 nm, or 561 nm) is deselected, and
then set the value to about 0.3% by adjusting the
power slider.
(3) Select the checkbox for the activation laser so that it
emits laser light for about a second, and then
immediately deselect it.
Check that the live image becomes bright once and
then returns to the previous state as a result of
activation causing the fluorescent probe to
fluoresce.
When using multiple activation lasers, check this for
each activation laser.
If activation cannot be checked
If emission of the activation laser does not make the live
image bright, this can be improved by replacing the
buffer solution of the specimen with a fresh solution.
3-(1)
3-(3)
3-(2)
4-(2)
4-(1)
4-(3)