Chapter 6 Assembly
3 Replacement of Consumable Materials
74
2
Replacing the Fuse
•
To avoid electrical shock or damage to the instrument, turn off the
power switch (turn to “
{
”) and unplug the power cord before replacing
the fuse.
•
Use the specified fuse. Using a different fuse may damage the
instrument or cause a fire hazard.
Specified Fuse: 250V 1A, time-lag, low-breaking type, 5x20 miniature
fuse
x2
•
Make sure that the contact of the fuse is not damaged before installing
a new fuse. If the contact is damaged, a malfunction or overheating
may result.
•
Attach the fuse to the fuse holder securely. If not, the fuse may come
loose or a contact failure may occur, resulting in overheating or smoke.
•
Put back the fuse holder securely to its original position.
(1) Turn off the power switch (turn to “
{
”) and unplug the power cord.
(2) Remove the fuse holder using a flatblade screwdriver.
(3) Remove the fuses and attach new fuses.
(4) Make sure that the voltage indication shown in the window of the fuse holder corresponds
to the voltage provided in your region. If not, change the voltage setting by relocating
the voltage selector. For details, see P.68.
(5) Put the fuse holder back in place.
WARNING
CAUTION
Fuses
Fuse Holder
Voltage Indication
Содержание eclipse E200
Страница 1: ...ECLIPSE E200 Instructions M313E 04 2 CF 2 ...
Страница 2: ......