Chapter 10 Technical Specifications
86
Model for 220, 230, 240 V Areas
•
Input voltage:
Select from 220 V, 230 V or 240 V AC by relocating
the fuse holder in the AC inlet.
•
Frequency: 50/60
Hz
•
Voltage fluctuation:
±
10%
•
Rated current:
0.4 A max.
•
Fuse rating:
250 V, 1A, time-lag low-breaking type, 5
×
20 miniature
fuse
×
2
•
Power Cord:
Use only the following power supply cord. Using the
wrong power cord could result in danger or fire.
The protection Class I equipment should be connected
to PE (protective earth) terminal.
•
Approved according to EU/EN standards,
3 conductor grounding Type H05VV-F, 3 m long
maximum, rated at 250 V AC minimum.
•
Protection Class:
Class I
(8) Operating
Environment
•
Room Temperature:
0° to 40°C
•
Relative Humidity:
85% Max. No condensation should form.
•
Altitude:
2000 m Max.
•
Pollution: Degree
2
•
Installation Category
(Overvoltage Category): Category 2
•
For indoor use only.
(9) Storage and Transport Environmental Conditions
•
Temperature: -20°
to
+60°C
•
Humidity:
max. 90%RH (non-condensing)
(10) Conforming Standards
•
The model for 100, 110 and 120 V areas is a UL-listed product.
•
The model for 220,230 and 240 V areas satisfies the EU IVD Directive (In vitro diagnostic medical
device directive).
(GM approved product : In vitro diagnostic medical device)
•
The model for 220, 230 and 240 V areas satisfies the EU LV Directive (Low Voltage Directive).
•
The model for 220, 230 and 240 V areas satisfies the EU EMC Directive. (EN 61326)
(The model for 100, 110 and 120 V areas is not covered by the FCC.)