Chapter 8 Technical Specifications
8.1
Specifications
80
Nikon Microscope ECLIPSE 55i
Model ECLIPSE
55i
Optical system
Infinity-corrected CF optical system
Objective: CFI60
Objectives:
Field number 22 (with ergonomic binocular tube/binocular
tube/cytodiagnostic unit), 25 (with trinocular eyepiece tube T/F)
Nosepiece: Sextuple
Focus up/down motion
Drive system: Manual coarse/fine motion
(calibration markings for fine motion: 1
µ
m/marking)
Stroke:
2 mm upward, 28 mm downward
With one-touch refocusing mechanism
Light source for
transmitted illumination
White LED
Input voltage
12V DC (provided from the AC adapter)
Power supply
AC adapter
Specified AC adapter
Manufacturer:
ILAN ELECTRONICS LTD.
Model: F1650K
Rated input voltage:
100-240V AC, 1.2A max. (when inputting 115V AC),
50/60Hz
Rated output voltage:
12V DC, 3.5A max.
Other:
UL-listed product, GS-approved, CE-certified
AC adapter power cord
•
When used in 100-120V areas:
UL-listed detachable power cord set (3 conductor grounding Type SVT, AWG 18,
3 m long maximum, rated at 125V AC minimum)
•
When used in 220-240V areas:
EU/EN-approved 3-conductor power cord set (3 conductor grounding Type
H05VV-F, 3 m long maximum, rated at 250V AC minimum)
Specified battery (Li-ion
rechargeable battery)
Manufacturer: Nikon
Code: EN-EL1
Rating: 7.4V
DC
Specified battery charger: MH-53 or similar
Operating conditions
Temperature: 0
to
40
°
C
Humidity:
85% RH max. (no condensation)
Altitude:
2000 m max.
Degree of pollution:
Degree 2
Installation: Category
II
Protection class:
Class I
Indoor use only
Содержание 50i
Страница 2: ...Microscope ECLIPSE 50i ECLIPSE 55i Instructions M317E 04 2 CF 2 1 2 ...
Страница 3: ......