27
INTRODUCTION: Before First Use
2.4.3
Charging the battery pack
Charge the battery pack before use.
The battery pack can be charged by either of the following methods:
•
Placing the main body on the station to charge (about 180 min.)
•
Inserting the battery pack into the battery slot of the station to charge (about 140 min.)
These charging methods can be performed at the same time if there are two battery packs.
It is recommended to prepare two or more battery packs in preparation for battery discharge, deterio-
ration, or such.
O
Charging the battery pack inserted in the main body
1
Connect the power cord.
“2.4.2 Power cord connection” (page 26)
2
Turn on
the power switch of the station.
3
With the battery pack inserted in the main
body, put it on the station as shown to the
right.
The main body battery charge indicator lights up and
charging starts.
The charge indicator goes out when charging is com-
plete.
If no battery pack is inserted in the main body
and the main body is placed on the station, the
charge indicator blinks.
CAUTION
• The battery pack has a limited lifetime. When the battery pack fails to hold a charge, replace it with a
new one.
• When any abnormality is found with the battery, the charge indicator blinks. Immediately turn off the
device to stop charging. Use a different battery pack.
• If any abnormalities such as odor, excessive heat, discoloration, or deformation are found during
use, charging, or storage of the battery pack, immediately turn off the device and stop using it. Con-
tact NIDEK or your authorized distributor.
• If charging is not indicated as complete after a specified period of time, or if the charge indicator
blinks, turn off the station and remove the battery pack.
Wait 5 minutes, insert the battery pack, then turn on the station to charge the battery pack.
If charging still does not indicate as complete, stop charging the battery pack and replace it with a
new one.
• Do not remove the battery pack while it is charging in the main body.