14
INTRODUCTION: Configuration
13
Forehead rest
Place against the patient’s forehead (above the eyebrows) to stabilize the main body.
Press to release the forehead rest.
14
Connection cable port
The cable connecting the main body and station is connected here.
When the cable is connected, the battery is not necessary for measurement. It also enables wired
LAN communication and printing from the station.
“2.5.1 Connection by cable” (page 36)
“3.11.1 Connecting peripheral devices” (page 82)
15
USB port
A USB flash drive is connected here.
Measurement data can be stored to the USB flash drive and copied to a computer.
CAUTION
• Equipment connected to the analog or digital interfaces must be certified according to the represen-
tative appropriate national standards (such as EN 60601-1 and IEC 60601-1). Furthermore, all con-
figurations shall comply with the system standard IEC 60601-1. Anyone who connects additional
equipment to the signal input part or signal output part configures a medical system, and is therefore
responsible that the system complies with the requirements of the system standard IEC 60601-1.
If in doubt, consult the technical service department of your local representative.
13
14
15
[Bottom view]