3

Unit model

DT-501X ● DT-501F

Display

Action mode

Digital tachometer mode

Flowmeter mode

Elapsed timecounter mode

Time width meter mode

Display 1

0 - 999999

6 digits

0:00:00 - 9:59:59 

(Hour:Minute:Second / base 60 display)

0:00:00 - 0:59:59 

(Hour:Minute:Second / base 60 display)

Display 2

-

0:00 - 999:99

(Hour:Minute:Second / base 10 display)

With zero suppression function

Decimal point position

0 to 5 digits after the decimal point(000000 ● 0.00000)

-

Number indicator

Red 7 segment LED, Letter height 22mm, 6 digits,  -  display available

LED lamp

8 (SIG, LL, L, H, HH, MAX, MIN, TEA)

Operation key

5 ●SET/SHIFT ,MEM/TEACH , ● /PARA , ● /FUNC ,ENT/MODE●

Input range

0.0067Hz ● 100kHz

10ms - 3600s

Measurement accuracy

±0.008● ±1digit

±0.1● ±1digit

Filter

Switches between 100kHz, 30kHz, 10kHz, and 0.02kHz using the parameter. 

Note that you can switch between only 10kHz and 0.02kHz in a magnetic sensor, and its contact is only 0.02kHz.

Display cycle

0.2, 0.5, 1, 2, 5, 10, 15, 30, 60 sec. (changeable in the parameter settings)

Dependent on the input signal

Pre-scale function

Parameter setting system using the front panel keys. The teaching (combination) of display values are also available.

Memory function

The maximum/minimum measurement values can be memorized and displayed in the indicator. 

(Switches the display using the MEM key)

Comparator function

The settings of the high set point 1, low set point 1, high set point 2, and low set point 2, and judgment results can 

be displayed on the LED lamp. The hysteresis setting of the high and low set point 1 values are also available.

Auto zero time

0.1 - 150 sec.

0.1 - 3600 sec.

Pre-arithmetic function

Updates the displayed value according to the elapsed time after the pulse stops.

Teaching function

Performs scaling automatically by setting the display value with a certain signal input. (only in the tachometer and flowmeter modes)

Insulation resistance

10MΩ or more (at DC500V Mega)

Voltage proof

AC1500V or more  1min

Operating temperature

0 - 45ºC(No condensation)

Operating humidity

35 - 85%RH (No condensation)

Operating atmosphere

No corrosive gas

Conforming standard

RoHS

Protection function

Front panel: IP66 (or equivalent), Rear terminal block: IP20

Casing material

ABS resin

External dimensions

● 96 × ● 48 × ● 92 ● (DIN)

Weight

Approx. 200g

FVT● FVC● TRT● BDC options : + approx. 50g

CPT options : + approx. 100g

●3. 

Specifications

●4. 

Component Part Names and Functions

●●●●●●●●●●●● ● ●●●●●● ●

●●●●●●●●●●●●●● ●●

Unit label

PS ℓ 

     h

cm 

   min

● 

     h

PS ℓ 

     h

cm 

   min

● 

     h

FVT
CPT

kHz rpm

ℓ 

   min

kHz rpm

ℓ 

   min

TRT
FVC

●:●:●

sec min

rps

Hz h:m:s sec min

rps

Hz

TRC

BCD

● : ● ● ●

m●● 

●● h

mm 

●● s

ℓ 

  ●s

r     

   min m:s:●  

     h

mm       

s

ℓ        

s

r      

   min

RMT

DRT

   min

km 

     h

mm       

   min

m       

   min

km 

     h

mm 

   min

SDT

SDC


10


10

21

Name

Function

● SIG lamp

Lights up when the sensor signal is input

● LL lamp

Lights up when the low set point 2 is judged.

● L lamp

Lights up when the low set point 1 is judged.

● H lamp

Lights up when the high set point 1 is judged.

● HH lamp

Lights up when the high set point 2 is judged.

● MAX amp

Lights up when the maximum value is displayed

● MIN lamp

Lights up when the minimum value is displayed

● TEA lamp

Lights up when the teaching function is set

● PEAK lamp

Not used

● BTM lamp

Not used

● Unit label space

Space for attaching the supplied unit labels

● Main display

Displays the measurement value

● SET/SHIFT key

Finishes the setting in various setting modes

Pressing this key with other keys switches to various setting modes

● MEM/TEACH key

Switches to the memory display

Pressing this key with the SET key goes to the teaching setting mode

● ●(UP)/PARA key

Changes the selected items in various setting modes, or numerical values

Pressing this key with the SET key switches to the parameter setting mode

● (NEXT)/FUNC key

Changes the selected digit in various setting modes

Pressing this key with the SET key switches to the function setting mode

● ENT/MODE key

Selects the changed item(s) in various setting modes

Pressing this key with the SET key switches to the mode setting mode

● Terminal block
● Rear panel
● Terminal block cover

21 Mounting adapter

*There are the input specifications in addition to the above specifications. For details, refer to p.5 - 6.

●Front

●Side

●Rear

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Содержание DT-501 FA-BCD

Страница 1: ...reading be sure to store this manual in a safe convenient place where operators can always refer to it easily Thank you for purchasing SHIMPO s Digital Tachometer DT 501X DT 501F For instructions on...

Страница 2: ...function settings p 17 19 Parameter settings p 8 15 Teaching function settings P7 1 Press the MEM key once to display the MAX value 2 Press the MEM key again to display the MIN value 3 Press the MEM...

Страница 3: ...densation Operating humidity 35 85 RH No condensation Operating atmosphere No corrosive gas Conforming standard RoHS Protection function Front panel IP66 or equivalent Rear terminal block IP20 Casing...

Страница 4: ...nt waterproofed but if water is splashed on the unit be sure to wipe it off the unit as soon as possible 1 Attach the provided waterproof gasket to the panel surface If waterproofing is not necessary...

Страница 5: ...d on the signal wire resulting in malfunctions Use shielded wire for input power connections with the shortest possible metal conduit Terminal block connection diagram PWR1 PWR2 E PU OC RE MG GND 12V...

Страница 6: ...nal connection Line driver Terminal block connection diagram Conduct the comparator function p 17 and memory function settings p 18 if necessary This unit requires no complicated calculation and imple...

Страница 7: ...speed is outside the input revolution speed range EE 2 is displayed and the teaching function cannot be used When the input frequency is outside the input frequency range EE 2 is displayed and the tea...

Страница 8: ...ted Circumference of the roller Revolution speed Normal display Mode setting Finish setting The main display shows 0 0 0 0 1 Press the SET key The display goes to mode 1 digital tachometer mode The nu...

Страница 9: ...000000 999999 0 F2 Low set point 1 value Sets the high set point 1 value 000000 999999 0 F3 High set point 2 value Sets the high set point 2 value 000000 999999 0 F4 Low set point 2 value Sets the low...

Страница 10: ...n Set each setting item for mode and parameter according to the following procedures For the setting method of functions refer to pages 16 and 17 As in the figure on the left bread passes through the...

Страница 11: ...tops in the elapsed timecounter mode the display will overflow after the specified time set with the auto zero function The overflow display does not mean there is a malfunction When the input signal...

Страница 12: ...em 0 00 00 0 00 0 00 1 100 sec 0 00 Select the 1 100 second display based on the operating time of the press machine P2 Measurement section 0 OFF 1 ON _ 1_ ON 0 Select 0 because the adjacent switch of...

Страница 13: ...e low set point 1 value Hour Minute Second display system 0 00 00 9 59 59 Second display system 0 00 Sets the low set point 1 value Second display system 0 00 999 99 F3 High set point 2 value Sets the...

Страница 14: ...ero time 0 1 150 sec _ _006 0 6 sec 6 sec Measure using the default setting value as no special instruction is provided P7 Input filter 10 30 100 0 02kHz _ 10_ 10kHz 10kHz Measure using the default se...

Страница 15: ...Parameter 2 setting Parameter 3 setting Parameter 4 setting Parameter 5 setting Parameter 6 setting Parameter 7 setting Finish setting The main display shows 0 00 Press the ENT key and start setting...

Страница 16: ...ost For the settings of the high set point 1 and 2 values as well as the low set point 1 and 2 values refer to Comparator Function on page 17 To the function settings The main display shows 0 0 0 1 F...

Страница 17: ...t Conditions under the low set point 1 and 2 value judgment Judgment conditions Set the high set point 1 low set point 1 high set point 2 and low set point 2 values using the function items F01 to F04...

Страница 18: ...y MAX MIN MAX MIN Normal measurement display Normal measurement display Normal measurement display Maximum value display Press the MEM key once during the normal measurement to display the maximum val...

Страница 19: ...E 3 E E 4 BTM MAX MINTEA PEAK BTM Normal display LED display test Key test Version display Mode display Test mode Press and hold the SHIFT MODE keys for 5 seconds or more Press the key to change the m...

Страница 20: ...er sensor blade cc etc Enter the capacity per blade that the sensor can read 0 0001 99999 1 0 P3 Scaling Unit coefficient value 0 00000 999999 1 P4 Decimal point display Designates the decimal point p...

Страница 21: ...99 0 F6 Judgment output timer at startup 2 Sets the time when the comparator judgment is output at startup 0 99 sec 0 sec F7 Minimum revolution speed Sets the revolution speed to be displayed as zero...

Страница 22: ...verage function is not set F8 setting value 0 When the auto zero function is activated the display becomes 0 in mode 2 and mode 3 Displays the measurement value per display cycle averaged by the frequ...

Страница 23: ...ect one out of these output options 11 12 13 14 15 16 17 18 19 20 Code Pin number Code C 4 20mA 1 19 C 4 20mA 2 20 NC 3 21 NC NC 4 22 NC NC 5 23 NC NC 6 24 NC NC 7 25 NC NC 8 26 NC V10 0 10V 9 27 V10...

Страница 24: ...log signal output offset Set the output offset value as the percentage of the output range for each analog signal __ 100 0 __100 0 0 4 20 mA 0 10 V 1 5 V 0 1 V Output range 16mA Output range 10V Outpu...

Страница 25: ...r PCS E36LMD Accessory side Plug PCS E36FS Cover PCS E36LA Both manufactured by HONDA TSUSHIN KOGYO CO LTD Positive logic negative logic available for BCD and decimal point output opted at Function 10...

Страница 26: ...output logic Set as 0 for negative logic set as 1 for positive logic decimal point output _ 0_ Negative logic _ 1_ Positive logic Negative logic F10 sets positive negative logic for BCD output Decima...

Страница 27: ...c open collector output Model CPT Output contact 1C Rated load Resistance load 250VAC 5A 100K operations 30VDC 5A 100K operations Induction load cos 0 4 250VAC 2 5A 100K operations 30VDC 2 5A 100K ope...

Страница 28: ...501XA BCD DT 501XD BCD BCD BCD output Differential input DT 501FA FVT FVT Voltage output DT 501FA FVT BCD BCD BCD output DT 501FA TRT TRT Transistor output DT 501FA TRT FVC FVC Voltage output DT 501FA...

Отзывы: