Table 11. General Connector Descriptions (Continued)
Connector
Description
Connector Type
REF
IN
Input connector for the Reference Clock.
MMPX (f)
OUT Output connector for the Reference Clock.
MMPX (f)
LO1
IN
Input connector for the external local oscillator
signal from the mmWave radio head, used for
the up and down conversion between IF and
mmWave.
MMPX (f)
OUT Output connector for the local oscillator signal
to the mmWave radio head, used for the up and
down conversion between IF and mmWave.
MMPX (f)
LO1 1 mmWave
Output connector that routes a local oscillator
signal to a secondary mmWave radio head.
SMA 27 GHz (f)
AUX 1
Reserved for internal use.
MMPX (f)
Table 12. LED Indicators
LED
Indications
ACCESS Indicates the basic hardware status of the module.
•
OFF
—The driver software has not yet loaded, or the module exceeded
approved operating temperature and thermal shutdown occurred or has
detected a problem with a power rail.
•
AMBER
—The module is being accessed.
Accessed
means that the device
setup registers are being written to in order to control the device.
•
GREEN
—The module is ready to be programmed by software.
ACTIVE •
OFF
—The module is idle.
•
RED
—The module detected an error state.
30
|
ni.com
|
PXIe-5831 Getting Started Guide