3.2 Display at power “ON” and initial settings
FRA5097
3-15
Table 3-1 List of initialization values (continued)
Setting items
Initialization
values
At
shipment
from
factory
At data
destruction
At
initialization
executed
Notes/Remarks
OTHERS
TITLE SET
(Cleared)
○
○
−
Also cleared at
re-measurement
and at power
supply switch on
BUZZER ON
○
○
○
DATE SET
YEAR
1988
−
○
−
MONTH
1
−
○
−
DAY
1
−
○
−
The actual
shipment date is
set at shipment
from factory
TIME SET
HOUT
0
−
○
−
MINITE
0
−
○
−
The actual
shipment time is
set at shipment
from factory
3.2.3
Warm-up
It takes more than 30 minutes after the power supply switching-on for the internal temperature of
the
FRA5097
to reach stable.
Perform measurement right after calibration is made, which shall be made after sufficient time of
warm-up has been made. Note that the measurement accuracy specification is met under the
condition immediately after calibration.
Conduct re-calibration when environment temperature has been changed.
Содержание FRA5097
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Страница 2: ......
Страница 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Страница 40: ......
Страница 122: ......
Страница 132: ......
Страница 156: ......
Страница 168: ......
Страница 188: ......
Страница 190: ......
Страница 191: ......
Страница 192: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN ...