9.7 Display
FRA5097
9-11
9.7 Display
●
Display device
6.5 inch color TFT LCD
●
Graph
Bode chart, Nyquist diagram, Nichols chart, Cole cole plot
(Cursor reading and autoscale available)
●
Graph display style
SINGLE/SPLIT
When SPLIT is chosen, the graph display area in the screen is split
into two, and two graphs can be displayed simultaneously
●
Measurement data display
Gain
Linear
±
9.999E+7 to
±
1.000E-8, as well as 0
Log
±999.999dB
Phase
-180.00 deg. to 179.99 deg.
0.00 deg. to 359.99 deg.
-360.00 deg. to -0.01 deg.
Enlarged
display
When the measurement mode is SINGLE or REPEAT,
measurement data are displayed enlarged in center of the LCD
display device
●
Auto scaling
A feature that automatically sets the graph display scale to optimum
Valid during measurements and also upon the initial data display
When the auto scale is OFF, the scale is fixed
●
Marker display
Normal marker
Displays a normal marker on the graph, and the data relevant to
the position specified by the normal marker are displayed on the
LCD numerically
Delta marker
Displays a normal marker and a delta marker on the graph, while
the difference between the normal marker and delta marker is
displayed numerically
●
Measurement data measurement condition display
Displays notable measurement conditions during the measurement of the data being displayed
●
Graph types
Bode chart
Amplitude - frequency, phase- frequency graph
Nyquist diagram
a+jb display
Cole cole plot
Displays the Nyquist diagram (a+jb) with the positive-negative of
the imaginary axis (b) being reversed
Nichols chart
The transfer function graph whose vertical axis is gain and
horizontal axis is phase
●
Condition setting/reference
By menu
●
Title display
Sets a title to measurement data and calculation data, and the title is displayed with the graph
●
Date and time display
Displays the date and time which are either up to date, or are those of the time the data were
sampled
Содержание FRA5097
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Страница 2: ......
Страница 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Страница 40: ......
Страница 122: ......
Страница 132: ......
Страница 156: ......
Страница 168: ......
Страница 188: ......
Страница 190: ......
Страница 191: ......
Страница 192: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN ...