Command Set
51
JA
start jog motion
USAGE
IMM
♦
MIP
♦
SYNTAX
xx
JA
nn or xxJA?
PARAMETERS
Description
xx
[ int ]
-
Controller number
nn
[ int ]
-
Jog setting
Range
xx
-
0
to
255
nn
-
0
to ±
8
Units
xx
-
None
nn
-
None
Defaults
xx
missing:
none
out of range:
none
nn
missing:
error 38, Parameter missing
out of range:
error 7, Parameter out of range
DESCRIPTION
This command will start a jog motion with the specified speed setting
indicated by
nn
.
nn = 0 Jog velocity = 0 full-steps/sec (0.00 mm/sec, for NSA25)
nn = ±1 Jog velocity = ±0.07 full-steps/sec (0.0005 mm/sec, for NSA25)
nn = ±2 Jog velocity = ±0.23 full-steps/sec (0.0015 mm/sec, for NSA25)
nn = ±3 Jog velocity = ±0.78 full-steps/sec (0.005 mm/sec, for NSA25)
nn= ±4 Jog velocity = ±3.12 full-steps/sec (0.02 mm/sec, for NSA25)
nn = ±5 Jog velocity = ±15.62 full-steps/sec (0.1 mm/sec, for NSA25)
nn = ±6 Jog velocity = ±48.87 full-steps/sec (0.3 mm/sec, for NSA25)
nn = ±7 Jog velocity = ±156.62 full-steps/sec (1.0 mm/sec, for NSA25)
RETURNS
If the “?” sign takes the place of
nn
value, this command reports the last
speed setting value.
REL.
COMMANDS
ST
-
Stop Motion
EXAMPLE
25JA-5
|
Start Jog motion with –50 fs/sec for controller 25
25JA?
|
Reads back the value of the last Jog speed from
controller 25
Содержание NewStep NSA12
Страница 1: ...Actuator and Motion Controller NewStep User s Manual TM...
Страница 20: ...20 Getting Started This page is intentionally left blank...
Страница 25: ...Operating NewStep 25 4 1 3 Single Actuator no Computer Interface minimal configuration...
Страница 26: ...26 Operating NewStep 4 1 4 Single Actuator with Computer Interface...
Страница 27: ...Operating NewStep 27 4 1 5 Two Actuators no Computer Interface...
Страница 28: ...28 Operating NewStep 4 1 6 Two Actuators with Computer Interface...
Страница 29: ...Operating NewStep 29 4 1 7 Multiple Actuators with Computer Interface...
Страница 32: ...32 Operating NewStep This page is intentionally left blank...