58
7.6
Model 350B Series Setup
Setting up the Model 350B Temperature Controller is a simple procedure.
An AC power cord, supplied with each unit, must be plugged into a properly grounded
outlet. The tilt up feet, at the bottom of the unit, may be extended to enhance the viewing
angle of the front panel. Connection to TE modules and temperature sensors are made
with one of the accessory cables (part #300-02 or #300-04).
7.6.1
Model 350B Operating Checklist
The following steps should be followed when operating a Model 350B Series
Temperature Controller.
a.
Check the AC voltage selection of the unit to be sure that it is compatible with
the outlet to be used. All units are factory preset for operation from 108–
132VAC.
b.
Connect the temperature sensor and TE module to the 15-pin output connector.
c.
Check that the proper temperature sensor bias is selected.
d.
Turn on the AC power switch and the unit will start-up with the output to the TE
module off. The display will be in the PRESET mode.
e.
Before setting any values on the DISPLAY select the operating mode, R/T or I
TE
.
f.
Change the display to the LIMIT display mode and set the maximum current
allowed to the TE module. For this, push the display button until the LIMIT SET
LED is lit. Turn the control knob until the display shows the desired limit current.
NOTE
The maximum current is specified by the manufacturer and must not be exceeded or
damage to the TE module may occur.
Содержание 350B
Страница 2: ...Temperature Controllers User s Manual Model 350B ...
Страница 7: ...vi This page is intentionally left blank ...
Страница 11: ...x This page is intentionally left blank ...
Страница 25: ...24 Getting Started This page is intentionally left blank ...
Страница 31: ...30 System Operation This page is intentionally left blank ...
Страница 61: ...60 ...