System Operation
29
29
4.2.4
AC Power Inlet
The input voltage setting is indicated in a small window on the face of the power module.
A small screwdriver is needed to flip down the panel once the AC line cord is removed.
Carefully remove plastic tumbler and reinsert it to show the appropriate power grid
voltage. The fuse is also located behind this panel and can be pulled out, for replacement
with the appropriate size, as indicated on the back panel.
All units are preset at the factory for operation at 108-132VAC, 60HZ.
The fuse must be changed for 198-242VAC and 216-264VAC operation.
Select:
100VAC for operation at 90-110VAC, 60Hz
120VAC for operation at 108-132VAC, 60Hz
220VAC for operation at 198-242VAC, 50Hz
240VAC for operation at 216-264VAC, 50Hz
The line cord supplied with each unit should be plugged only into a properly grounded
three-prong outlet to prevent electrical shock in the event of an internal short circuit to
the metal cabinet.
4.2.5
Frame or Chassis Terminal
This terminal provides the means to connect to the enclosure.
4.2.6
Power Inlet Socket
Plug the included power cord into the Power Inlet Socket on the rear of the instrument,
then plug the power cord into a wall socket with proper earth grounding.