Computer Interfacing
35
*SAV
Description
Save command.
Syntax
*SAV value
Argument
Value
Description
Value
0
Invalid
1
Saves current settings to working settings
2
Saves current settings to user settings
Remarks
The save command stores the current state of the instrument in non-volatile local
memory. This state is then recalled by using the
*RCL
recall command.
See Also
*RCL
*STB?
Description
Status Byte Register query.
Syntax
*STB?
Remarks
The Read Status Back query allows the programmer to read the Status Byte Register.
Response
Description
Status Byte Register
bit 0
Reserved
bit 1
Reserved
bit 2
Reserved
bit 3
Reserved
bit 4
Message Available
bit 5
Reserved
bit 6
Reserved
bit 7
Error Message Available
ADDRess
Description
USB address command.
Syntax
ADDRess
value
Remarks
The
ADDRess
command sets the instrument USB address. After changing USB address, the
communication with the instrument has to be re-initialized. This can be accomplished by calling
“InitSystem” function in the DLL available in the CD provided with the instrument.
Argument
Value
Description
Value
0
Reserved
1 to 99
Valid USB address range
See Also
ADDRess?
Содержание 350B
Страница 2: ...Temperature Controllers User s Manual Model 350B ...
Страница 7: ...vi This page is intentionally left blank ...
Страница 11: ...x This page is intentionally left blank ...
Страница 25: ...24 Getting Started This page is intentionally left blank ...
Страница 31: ...30 System Operation This page is intentionally left blank ...
Страница 61: ...60 ...