25
5.7
Distance accuracy correction
If the System produces any measurement errors that lie
beyond unit specification, it may be an indication that unit
requires calibration similar to one conducted at
production level. To do accuracy correction, please select
Corr
from
CAL
menu.
Perform correction according to required measurement
parameters by pressing Action button, see table below:
Active mode of
measurement
(
Std
)
Correction
parameter
Condition
d
Distance
Install unit at 1
meter
1
from flat
target
2
E
Elevation
Put unit in a level
plane
1 – Accuracy of distance correction will depend on
accuracy of installation against target. If distance is
0.9 meters, then unit will have 0.1 m offset during
all future measurements.