![NEW BRUNSWICK SCIENTIFIC CO-150 Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/new-brunswick-scientific/co-150/co-150_manual-to-operations_1677085030.webp)
30
M1310-0050 Model CO-150 CO
2
Incubator
User’s Guide
The outer door is kept closed to minimize heat loss through the glass door.
Remember to re-set the CO2 set point to the desired level after sampling.
We recommend that you perform a CO
2
AutoZero prior to sampling (
see Section
3.5
).
3. We also recommend that you AutoZero the CO
2
system once a month to ensure
that the CO
2
level is correct.