Programming Instructions
79
RNS
®
System User Manual
•
Suggested initial responsive therapy settings:
•
In general, stimulation should be delivered to the leads and electrodes from which the activity
of interest is observed.
•
Program
B
URST
1
for electrodes on one lead. Program
B
URST
2
for electrodes on the other
lead.
•
Program
C
URRENT
to 1.0mA.
•
Other parameters should be left at their default values:
•
F
REQUENCY
= 200 Hz
•
PW
PER
P
HASE
= 160 µs
•
B
URST
D
URATION
= 100 ms
•
Copy
T
HERAPY
1
settings to
T
HERAPY
2
through
T
HERAPY
5.
•
Program settings into the neurostimulator on the
P
ROGRAM
S
ETTINGS
tab.
Figure 37: Therapy 1 tab.