Responsive Neurostimulator System
12
RNS
®
System User Manual
WARNING: I
NTERACTION WITH
I
MPLANTED
C
ARDIAC
D
EVICES
Possible effects of implanted device interaction with an implanted cardiac device (e.g.,
pacemaker or defibrillator) include the following:
•
Defibrillation therapy from an implanted defibrillator may damage the device.
•
The electrical pulses from the neurostimulation system may interact with the
sensing operation from a cardiac device and could result in an inappropriate
response of the cardiac device and vice versa.
W
ARNINGS
–
P
ROGRAMMER
WARNING: P
OTENTIAL
S
HOCK
Submerging any part of the programmer, or operating the programmer in or near a wet
environment, may result in an electrical shock.
The programmer must be disconnected from the electrical outlet prior to cleaning to
avoid the potential of electrical shock.
Electrical shock may occur if the programmer AC adapter and power cord are not
properly connected to a grounded power source.
Do not attempt to modify or repair the wand or programmer. Contact NeuroPace
Customer Support for assistance.
WARNING: N
EURO
P
ACE COMPONENTS
Use of accessories, transducers, and cables other than those provided by NeuroPace
could result in increased electromagnetic emissions or decreased electromagnetic
immunity of the RNS® System and result in improper operation. Do not plug the wand
into equipment other than the programmer because it could damage the wand. Do not
use a USB cable extension from the programmer to the wand.
WARNING: P
ORTABLE
R
ADIO
F
REQUENCY
(RF)
C
OMMUNICATIONS
E
QUIPMENT
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 12 inches (30 cm) to any part of
the RNS® System, including cables. Otherwise, degradation of the performance of the
RNS® System could result.
WARNING: N
EURO
P
ACE
®
E
QUIPMENT
P
LACEMENT
Use of NeuroPace
®
equipment (for example, remote monitor or programmer) adjacent
to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, the NeuroPace
®
equipment and other
equipment should be observed to verify that they are operating normally.
W
ARNINGS
–
M
EDICAL
E
NVIRONMENT
WARNING: L
ITHOTRIPSY
The effects of extracorporeal shock wave lithotripsy on the RNS
®
System have not
been studied. Exposure to high-output ultrasonic frequencies may damage the RNS
®