NOTE: UT is the voltage of the alternative network before application of the test level.
Directives and manufacturer’s declaration – Electromagnetic immunity
The Saphyr is designed to be used in the electromagnetic environment specified below. The Saphyr user should make sure
that it is used in such an environment.
Immunity test
Test level
according to
CEI 60601
Level of
conformity
Electromagnetic environment – Directives
Radiated RF
CEI 61000-4-3
±3 V/m
from 80 MHz
up to 2500 MHz
3V/m AM
80% - 1 kHz
80 to
2500 MHz
RF communication mobile and portable equipment should
not be used too close to any part of the Saphyr, including the
cables. The recommended separation distance calculated from
the applicable equation to the transmitter frequency should be
adhered to.
Recommended separation distance :
d= 1.2√P
d= 1.2√P from 80 MHz to 800 MHz
d= 1.2√P from 800 MHz to 2500 MHz
where P is the maximum transmitter exit power in watts (W)
depending on the make of transmitter and the recommended
separation distance in metres (m).
The field intensities of the fixed RF transmitters determined
by on-site electromagnetic investigation
a
should be below the
level of conformity in each frequency range
b
. Interference can
occur close to equipment marked with the following symbol:
NOTE 1: At 80 MHz and at 800 MHz the highest frequency range applies.
NOTE 2: These directives may not apply in all situations. Electromagnetic propagation is modified by absorption and
by reflections from structures, objects and people.
The field strength of fixed transmitters such as base stations for radiotelephones (cell phones or wireless) and terrestrial
mobile radios, amateur radio, AM and FM broadcasting and television broadcasting cannot be theoretically predicted with
precision. To evaluate the electromagnetic environment resulting from fixed RF, an on-site electromagnetic investigation
should be considered. If the strength of the field measured at the place where the Saphyr is used exceeds the level of
RF conformity applicable above, the Saphyr should be monitored to check that it is functioning normally. If unusual
performance is observed, additional measures may be necessary, such as reorienting or repositioning the Saphyr.
Over the range of frequencies from 150 KHz to 80 MHz, the field strength should be less than 3V/m.
a -
b -
MU SAPHIR.ZONE 3.indd 19
18/05/2010 18:06:32
Содержание SAPHYR CX
Страница 23: ...21 MU SAPHIR ZONE 3 indd 21 18 05 2010 18 06 32...
Страница 24: ...22 MU SAPHIR ZONE 3 indd 22 18 05 2010 18 06 32...
Страница 28: ...26 Neurelec Neurelec Neurelec 1 2 3 4 20 5 Saphyr MU SAPHIR ZONE 3 indd 26 18 05 2010 18 06 33...
Страница 31: ...29 Saphyr CX Saphyr CX SaphyrCX K1245 K XXXX MU SAPHIR ZONE 3 indd 29 18 05 2010 18 06 33...
Страница 34: ...32 4 P1 P2 P3 P4 0 Saphyr 5 Saphyr P1 P2 P3 P4 FM P1 P2 MU SAPHIR ZONE 3 indd 32 18 05 2010 18 06 33...
Страница 35: ...33 6 7 Saphyr 2 1 1 Neurelec 1 Saphyr hi fi Saphyr P1 P2 P3 P4 MU SAPHIR ZONE 3 indd 33 18 05 2010 18 06 34...
Страница 37: ...35 Saphyr 2 20 C 2002 96 CEE DEEE 2 Saphyr NF EN 606601 1 2 2007 MU SAPHIR ZONE 3 indd 35 18 05 2010 18 06 34...
Страница 38: ...36 5 Saphyr 1 Saphyr 11 9 Saphyr MU SAPHIR ZONE 3 indd 36 18 05 2010 18 06 35...
Страница 40: ...38 2 1 2 3 12 4 Neurelec 1 2 1 2 4 Saphyr Neurelec MU SAPHIR ZONE 3 indd 38 18 05 2010 18 06 35...
Страница 45: ...43 MU SAPHIR ZONE 3 indd 43 18 05 2010 18 06 36...
Страница 46: ...44 MU SAPHIR ZONE 3 indd 44 18 05 2010 18 06 36...
Страница 67: ...65 MU SAPHIR ZONE 3 indd 65 18 05 2010 18 06 39...
Страница 68: ...66 MU SAPHIR ZONE 3 indd 66 18 05 2010 18 06 39...
Страница 89: ...87 MU SAPHIR ZONE 3 indd 87 18 05 2010 18 06 42...
Страница 90: ...88 MU SAPHIR ZONE 3 indd 88 18 05 2010 18 06 42...
Страница 111: ...109 MU SAPHIR ZONE 3 indd 109 18 05 2010 18 06 46...
Страница 112: ...110 MU SAPHIR ZONE 3 indd 110 18 05 2010 18 06 46...