Data Report
After power on, the device will immediately send a version packet report. Then, it will send another report with the data of pH
value and temperature of the solution
after it is
powered on for 20s.
The device sends data according to the default configuration before any other configuring.
Default setting:
MaxTime: Max Interval = 3min = 180s
MinTime: The MinTime configuration is not available.
*But the software has restriction, MinTime must be configured a number greater than 0.
Note:
Report interval is based on the factory default.
R718PA8 reports the pH value and temperature of the solution.
The device reported data parsing please refer to Netvox LoraWAN Application Command document and Netvox Lora Command
http://loraresolver.netvoxcloud.com:8888/page/index
5.1 Example of ReportDataCmd
FPort
:
0x06
Bytes
1
1
1
Var(Fix=8 Bytes)
Version
DeviceType
ReportType
NetvoxPayLoadData
Version
– 1 byte –0x01
——
the Version of NetvoxLoRaWAN Application Command Version
DeviceType
– 1 byte – Device Type of Device
The devicetype is listed in Netvox LoRaWAN Application Devicetype doc
ReportType
– 1 byte –the presentation of the NetvoxPayLoadData
,
according the devicetype
NetvoxPayLoadData
– Fixed bytes (Fixed =8bytes)
Device
Device
Type
Report
Type
NetvoxPayLoadData
R718PA Series
(R718PA8)
0x57
0x08
Battery
(1Byte, unit:0.1V)
PH
(2Bytes ,0.01pH)
TemperaturewithPH
(Signed 2Bytes,unit:0.01°C)
ORP
(Signed 2Bytes ,1mv)
Reserved
(1Byte,fixed 0x00)
Uplink: 015708000226092EFFFF00