UPM 2000
6.
Remove the calibration check weight. It is normal for the unit to
take 30 seconds to 1 minute to completely return to zero.
7.
If the reading is 98.3 to 101.8 grams, the unit is in good working
order. If the unit is outside of this range, the unit should be
returned for repair and/or calibration.
Maintenance
Maintenance of the UPM 2000 power meter is straightforward, requiring little more
than keeping it clean. It is important to keep dust and dirt out of the transducer tank.
Rinsing the tank with clean water before use, along with proper storage in the
carrying case, is all that is required. It is also important that the cone and hanger be
removed and stored in the carrying case after use.
Do not use
light detergent and non-abrasive cleaning pad is acceptable on the exterior of the
unit.
Service
The mechanical assembly of the UPM 2000 contains no parts that can be serviced
by the user.
The unit should be returned to Netech Biomedical for repair or
calibration.
The alignment and adjustment parameters are critical to the
performance of the UPM 2000 and can be performed only at the factory or
designated distributor.
For service, pack the UPM 2000 according to the
Repackaging and Shipping
instructions
in this chapter.
Failure
Call Netech Customer Service at 1-800-547-6547 (U.S.) or 1-631-531-0100
(International) for shipping instructions.
any type
of cleaner in the tank or on the transducer cone. The use of a
Netech Corporation
110 Toledo St, Farmingdale, NY 11735
http://www.Netech.org
41
1-800-547-6547
to do so may void warranty.
Содержание UPM 2000
Страница 1: ...UPM 2000 Ultrasound Power Meter User Manual Manual P N 1200 User Manual Initial Release 3 12 2008...
Страница 2: ...UPM 2000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www Netech org 2 1 800 547 6547...
Страница 12: ...UPM 2000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www Netech org 12 1 800 547 6547...
Страница 38: ...UPM 2000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www Netech org 38 1 800 547 6547...
Страница 46: ...UPM 2000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www Netech org 46 1 800 547 6547...