Using
lding the
the Universal Transducer Clamp Assembly and Centering Rings
The universal transducer clamp assembly frees the operator from ho
transducer treatment (radiating) head in the coupling water and prevents transducer
movement, ensuring more accurate and stable readings. The centering rings help to
properly center and aim the transducer head in the tank.
1. Screw the clamp support post on the screw stud located at the rear of
the unit. Securely tighten by
hand
.
Do not use pliers to tighten post.
2. Center the transducer clamp assembly over the transducer well, as
shown, using the hook clamp to tighten it into place.
1-800-547-6547
3. Place a centering ring, with the hole sized to fit the transducer, on
top of the well. The centering ring helps assure that the sound head
is properly centered and aimed vertically toward the transducer
cone. If there
ole size, one
can be cut to t
ged hole is
centered in the
4. Loosen the side screw of the clamp assembly so that clamp rotation is easy,
and roughly c
ucer well for vertical
transducers w
n it such that it aligns
with the transducer handle (as shown in Figure B below).
are no centering rings with the proper h
he proper size. Just be sure that the enlar
ring.
enter the clamp opening over the transd
ithout handles (as shown in A), or positio
Hook Clamp
Содержание UPM 2000
Страница 1: ...UPM 2000 Ultrasound Power Meter User Manual Manual P N 1200 User Manual Initial Release 3 12 2008...
Страница 2: ...UPM 2000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www Netech org 2 1 800 547 6547...
Страница 12: ...UPM 2000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www Netech org 12 1 800 547 6547...
Страница 38: ...UPM 2000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www Netech org 38 1 800 547 6547...
Страница 46: ...UPM 2000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www Netech org 46 1 800 547 6547...