QRS
Ventricular Tachycardia
Rate
Amp
Width
Rate
Amp
(BPM) (mV)
(ms) (BPM) (mV)
80
1
100
195
2.0
80
1
100
195
1.0
80
1
100
195
4.0
RESPIRATION:
Select Resp in the Simulator Menu to access the Respiration selections.
Resp Menu
Rate
Imp
dR
Apne
Respiration waveforms are generated with four selectable rates, baseline impedances, and delta
impedance variations.
The Respiration default settings are:
Respiration Rate
30 BPM
Impedance
500
Ohms
Delta Impedance 1.0 Ohm
Apnea
Off
The value changes that may be made are:
Rate: 15, 30, 60, 120 BPM.
Impedance: 250, 500, 750, and 1000 Ohms.
Delta Impedance: 0.1, 0.5, 1.0, and 1.5 Ohms.
Apnea:
Resp/Apne=Off
Off Cont 12s 32s
Off:
Apnea is absent. Normal respiration waveforms are generated.
Cont:
Continuous Apnea. No respiration waveforms are generated.
12s:
No respiration waveform is generated for 12 seconds.
32s:
No respiration waveform is generated for 32 seconds.
BLOOD PRESSURE:
Select Bp in the Simulator Menu to access Blood Pressure selections.
Two blood pressure waveforms are generated with selections of 12 static and 6 dynamic values.
BP1 values are selected from the listed choices and BP2 values are ½ of those selected for BP1.
BP1 Menu
Dyna Stat
The Dynamic pressure waveforms track the Normal Sinus Rhythm rates.
The default settings for pressure values are:
BP1 Dynamic: 120/80 Static:
0
BP2 Dynamic: 60/40 Static:
0
The Dynamic pressure value selections are:
100/60 120/80
50/10
60/20
70/30
80/40
The Static pressure value selections are:
0 5 10 20
25 30 40 50
100 150 200 300
Minisim 1000 User Manual
16