GENERAL 

OVERVIEW 

       

 

 

1.1 Introduction 

 

GENERAL DESCRIPTION 

 

The MiniSim 1000 Multi-Parameter Patient Simulator is an advanced microcontroller 
based instrument.  It is designed to simulate patient signals of ECG, arrhythmia, invasive 
blood pressure, respiration, and temperature.  The device also simulates square, sine, 
triangle, and pulse performance waveforms. 
 
The easy to operate MiniSim 1000 is menu driven via eight tactile feel soft keys.  All 
functions are displayed on a two line sixteen character LCD display.  
 
The small hand held instrument is powered by one 9 Volt battery or an optional AC 
adapter. The MiniSim 1000 is CE marked and shipped with a Certificate of Calibration 
traceable to the NIST.   

 

The MiniSim 1000 is a rugged instrument that performs its simulations quickly, 
accurately, and with ease. 

 

1.2 Specifications 

 

ECG:  

12 leads with independent outputs referenced to RL. 

 

NORMAL SINUS RHYTHM (NSR) 
ECG Rates: 

30, 60, 70, 80, 90, 100, 120, 150, 180, 210, 240, 270, 300, and 350 BPM. 

Accuracy  0.5%. 

 
Amplitudes: 

0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II. Lead 1 is 0.6 X 

Lead II, Lead III is 0.4 X Lead II, and Lead V is the same as Lead II. 

 
High Level: 

500 X low level output on Lead II.  

Accuracy: 2% (1-5mV). 
 

PERFORMANCE WAVEFORMS (PERF) 
Sine, Square, Triangle, Pulse 
 
Frequencies: 

0.1 to 0.9 in 0.1 Hz steps.                             

1.0    to 9.0 in 1.0 Hz steps.  
  10   to 100 in 10 Hz steps. 
 

Accuracy

: 1% 

 

Amplitudes: 

0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II. 

Pulse: 

20 ms pulse of 1mV amplitude repeated at 4 second intervals. 

 
R WAVE DETECTION (RWD) 

Minisim 1000 User Manual

 

5

 

 

Содержание MiniSim 1000

Страница 1: ...MiniSim 1000 Multi Parameter Patient Simulator INSTRUCTION MANUAL 110 Toledo Street Farmingdale N Y 11735 Phone 631 531 0100 Fax 631 531 0101 www Netech org Minisim user manual...

Страница 2: ...a product that proves to be defective provided you return the product shipping prepaid to Netech Corporation This warranty does not apply if the product has been damaged by accident or misuse or as th...

Страница 3: ...rademarks of Netech Corporation Any other trademark names used in this manual are only for editorial purposes and the benefit of the respective trademark owner with no intention of improperly using th...

Страница 4: ...ies 10 2 Operating Instructions Controls and Indicators 08 Baseline ECG Waveform 09 Performance Waveforms 11 Arrhythmia Waveforms 12 Automatic Testing Sequences 13 Respiration Waveforms Apnea 16 Blood...

Страница 5: ...libration traceable to the NIST The MiniSim 1000 is a rugged instrument that performs its simulations quickly accurately and with ease 1 2 Specifications ECG 12 leads with independent outputs referenc...

Страница 6: ...60 and 120 BPM Accuracy 1 Apnea Off continuous 12 seconds and 32 seconds Lead Configuration Leads I and II BLOOD PRESSURE Impedance 350 Ohms Excitation 2 to 16 Volts Sensitivity 5 V V mmHg Static 0 5...

Страница 7: ...3 ACCESSORIES Description Part Number User Manual 300 Manual Pressure Cable Open Ended 303 Temperature Cable Open Ended 304 OPTIONAL ACCESSORIES Description Part Number Soft Carrying Case 301 110 VAC...

Страница 8: ...rature Default Setting YSI 700 series set to 25 C YSI 400 series set to 25 C CONTROLS AND INDICATORS 1 On Off Key 2 ECG Lead Snaps 3 LCD Display 4 Temperature Connector 5 Blood Pressure Connector 6 Hi...

Страница 9: ...ameter for simulation ECG Resp BP or Temp 4 Continue to select choices under the parameter to be simulated until the functional characteristics have been chosen 5 Each Parameter has default settings E...

Страница 10: ...with pacer waves with a 0 1 ms or 2 ms width and amplitude of 2 mV at 80 BPM Note 0 1 ms pacer width for AP and ASP can only be selected by going to PCR A selection other than 0 1 ms produces 2 ms wi...

Страница 11: ...orm with the Q wave modified as in Infarction and the ST segment elevated 7 as in ST Elevation Selecting TalT branches to TalT Rate 80 Rate Ampl An 80 BPM QRS test signal of 1 mV amplitude and 100 ms...

Страница 12: ...at Normal Sinus Rhythm is generated at 80 BPM with every 10th beat missing AFLT Atrial Flutter Varying ECG rates with 12 cycles at 60 BPM for 12 seconds 9 cycles at 90 BPM for 6 seconds 15 cycles at 1...

Страница 13: ...R is generated at 80 BPM Each time the F4 function key is pressed 12 PVCs are generated at different intervals PV24 Premature Ventricular Contraction 24 NSR is generated at 80 BPM Each time the F1 fun...

Страница 14: ...equence Amplitude Width Rate Time mV ms BPM sec 0 5 100 80 30 210 20 70 80 30 210 20 120 80 30 210 20 2 0 100 80 30 210 20 70 80 30 210 20 120 80 30 210 20 5 0 100 80 30 210 20 70 80 30 210 20 120 80...

Страница 15: ...cept for the QRS rate that becomes 30 BPM and the pacer rate that becomes 80 BPM During the ASP and AP test sequences the display will show the pacer amplitude the pacer width and the QRS rate For pac...

Страница 16: ...ormal respiration waveforms are generated Cont Continuous Apnea No respiration waveforms are generated 12s No respiration waveform is generated for 12 seconds 32s No respiration waveform is generated...

Страница 17: ...nd YSI 700 CALIBRATION The MiniSim 1000 is shipped from the factory fully calibrated with a Certificate of Calibration traceable to the NIST Annual calibration is recommended Contact the Customer Serv...

Страница 18: ...Appendix A Minisim 1000 User Manual 18...

Отзывы: