The automatic sequence is initiated when the test is selected and continues until all of the test
patterns have been generated. During the test sequence the values generated are displayed on
the LCD. At the completion of the automatic test, the display will return to its steady mode.
RWD:
R Wave Detection.
R waves are generated with three varying parameters of amplitude,
width, and rate. The three parameter values are displayed on the LCD while generated. Each test
waveform is displayed for 20 seconds.
The complete waveform test sequence is performed in three separate sets.
Set 1: Variable Parameters
Amplitude: 0.5, 2, 5 mV
Width: 100, 70, 120 ms
Rate:
80, 30, 210 BPM
Set 1 Complete Test Sequence
Amplitude Width Rate Time
(mV) (ms) (BPM) (sec)
0.5
100
80, 30, 210 20
70
80, 30, 210 20
120
80, 30, 210 20
2.0
100
80, 30, 210 20
70
80, 30, 210 20
120
80, 30, 210 20
5.0
100
80, 30, 210 20
70
80, 30, 210 20
120
80, 30, 210 20
As the test progresses, the indicated heart rate displayed on the patient monitor should be within
±
10% or
±
5 BPM whichever is greater of the applied rate.
Set 2: Variable Parameters
Amplitude: 0.15 mV
Width: 70, 120 ms
Rate: 30, 210 BPM
Set 2 Complete Test Sequence
Amplitude Width Rate Time
(mV) (ms) (BPM) (sec)
0.15 70 30, 210 20
0.15 120 30, 210 20
The monitor will not respond to the waveforms in this sequence Set.
Set 3: Variable Parameters
Amplitude: 1.0 mV
Width: 10 ms
Rate: 30, 210 BPM
Set 3 Complete Test Sequence
Amplitude Width Rate Time
(mV) (ms) (BPM) (sec)
1.0 10 30, 210 20
The patient monitor will not respond to the waveforms in the Set 3 test sequence.
Minisim 1000 User Manual
14