The automatic sequence is initiated when the test is selected and continues until all of the test 
patterns have been generated.  During the test sequence the values generated are displayed on 
the LCD. At the completion of the automatic test, the display will return to its steady mode.

 

 
RWD:

 

R Wave Detection. 

R waves are generated with three varying parameters of amplitude, 

width, and rate.  The three parameter values are displayed on the LCD while generated.  Each test 
waveform is displayed for 20 seconds. 
 
The complete waveform test sequence is performed in three separate sets. 
 
Set 1: Variable Parameters 
 

Amplitude: 0.5, 2, 5 mV 

 

Width:       100, 70, 120 ms 

 

Rate: 

     80, 30, 210 BPM 

 
Set 1 Complete Test Sequence 
 
   Amplitude    Width               Rate         Time 
       (mV)         (ms)               (BPM)       (sec) 
        0.5 

   100 

80, 30, 210           20 

 

                   70 

80, 30, 210           20 

                         120 

80, 30, 210           20 

 
        2.0 

   100 

80, 30, 210           20 

 

                   70 

80, 30, 210           20 

           

   120 

80, 30, 210           20 

 
         5.0 

   100 

80, 30, 210           20 

 

                   70 

80, 30, 210           20 

           

   120 

80, 30, 210           20 

 
As the test progresses, the indicated heart rate displayed on the patient monitor should be within  

±

 10% or 

±

 5 BPM whichever is greater of the applied rate. 

 
Set 2: Variable Parameters 
 

Amplitude: 0.15 mV 

 

Width:        70, 120 ms 

 

Rate:          30, 210 BPM 

 
Set 2 Complete Test Sequence 
 
   Amplitude    Width         Rate        Time 
       (mV)         (ms)         (BPM)       (sec) 
        0.15           70         30, 210      20 
        0.15          120        30, 210      20 
 
The monitor will not respond to the waveforms in this sequence Set. 
 
Set 3: Variable Parameters 
 

Amplitude: 1.0 mV 

 

Width:        10 ms 

 

Rate:          30, 210 BPM 

 
 
 
 
Set 3 Complete Test Sequence 
 
   Amplitude    Width         Rate        Time 
       (mV)         (ms)        (BPM)       (sec) 
        1.0            10           30, 210      20 
The patient monitor will not respond to the waveforms in the Set  3 test sequence. 

Minisim 1000 User Manual

 14

 

 

Содержание MiniSim 1000

Страница 1: ...MiniSim 1000 Multi Parameter Patient Simulator INSTRUCTION MANUAL 110 Toledo Street Farmingdale N Y 11735 Phone 631 531 0100 Fax 631 531 0101 www Netech org Minisim user manual...

Страница 2: ...a product that proves to be defective provided you return the product shipping prepaid to Netech Corporation This warranty does not apply if the product has been damaged by accident or misuse or as th...

Страница 3: ...rademarks of Netech Corporation Any other trademark names used in this manual are only for editorial purposes and the benefit of the respective trademark owner with no intention of improperly using th...

Страница 4: ...ies 10 2 Operating Instructions Controls and Indicators 08 Baseline ECG Waveform 09 Performance Waveforms 11 Arrhythmia Waveforms 12 Automatic Testing Sequences 13 Respiration Waveforms Apnea 16 Blood...

Страница 5: ...libration traceable to the NIST The MiniSim 1000 is a rugged instrument that performs its simulations quickly accurately and with ease 1 2 Specifications ECG 12 leads with independent outputs referenc...

Страница 6: ...60 and 120 BPM Accuracy 1 Apnea Off continuous 12 seconds and 32 seconds Lead Configuration Leads I and II BLOOD PRESSURE Impedance 350 Ohms Excitation 2 to 16 Volts Sensitivity 5 V V mmHg Static 0 5...

Страница 7: ...3 ACCESSORIES Description Part Number User Manual 300 Manual Pressure Cable Open Ended 303 Temperature Cable Open Ended 304 OPTIONAL ACCESSORIES Description Part Number Soft Carrying Case 301 110 VAC...

Страница 8: ...rature Default Setting YSI 700 series set to 25 C YSI 400 series set to 25 C CONTROLS AND INDICATORS 1 On Off Key 2 ECG Lead Snaps 3 LCD Display 4 Temperature Connector 5 Blood Pressure Connector 6 Hi...

Страница 9: ...ameter for simulation ECG Resp BP or Temp 4 Continue to select choices under the parameter to be simulated until the functional characteristics have been chosen 5 Each Parameter has default settings E...

Страница 10: ...with pacer waves with a 0 1 ms or 2 ms width and amplitude of 2 mV at 80 BPM Note 0 1 ms pacer width for AP and ASP can only be selected by going to PCR A selection other than 0 1 ms produces 2 ms wi...

Страница 11: ...orm with the Q wave modified as in Infarction and the ST segment elevated 7 as in ST Elevation Selecting TalT branches to TalT Rate 80 Rate Ampl An 80 BPM QRS test signal of 1 mV amplitude and 100 ms...

Страница 12: ...at Normal Sinus Rhythm is generated at 80 BPM with every 10th beat missing AFLT Atrial Flutter Varying ECG rates with 12 cycles at 60 BPM for 12 seconds 9 cycles at 90 BPM for 6 seconds 15 cycles at 1...

Страница 13: ...R is generated at 80 BPM Each time the F4 function key is pressed 12 PVCs are generated at different intervals PV24 Premature Ventricular Contraction 24 NSR is generated at 80 BPM Each time the F1 fun...

Страница 14: ...equence Amplitude Width Rate Time mV ms BPM sec 0 5 100 80 30 210 20 70 80 30 210 20 120 80 30 210 20 2 0 100 80 30 210 20 70 80 30 210 20 120 80 30 210 20 5 0 100 80 30 210 20 70 80 30 210 20 120 80...

Страница 15: ...cept for the QRS rate that becomes 30 BPM and the pacer rate that becomes 80 BPM During the ASP and AP test sequences the display will show the pacer amplitude the pacer width and the QRS rate For pac...

Страница 16: ...ormal respiration waveforms are generated Cont Continuous Apnea No respiration waveforms are generated 12s No respiration waveform is generated for 12 seconds 32s No respiration waveform is generated...

Страница 17: ...nd YSI 700 CALIBRATION The MiniSim 1000 is shipped from the factory fully calibrated with a Certificate of Calibration traceable to the NIST Annual calibration is recommended Contact the Customer Serv...

Страница 18: ...Appendix A Minisim 1000 User Manual 18...

Отзывы: