14
CAUTION
When switching from Normal to Reverse or vice
versa, make sure to pause the rocker switch in the
off (middle position). Power to the outlet will be
OFF in the open neutral position.
3.
DEVICE CURRENT MEASUREMENT:
Move the Mode switch to DEVICE CURRENT. Set
the polarity switch to normal. Turn the DUT On.
The LCD will display the Device Current in Amps.
CAUTION
Do not attempt to open the unit. There are no
user serviceable parts inside. Further, the
warranty will be void if the unit is opened by other
than Netech trained personnel.
PERFORMANCE CHECK
1.
Connect the LKG 601 to a live 110 VAC (or 220
VAC for the LKG 601-220) outlet. Display will
turn on and read 0
±
1 when the Mode switch is
in the Case Leakage position. In the CORD
RESISTANCE position the display will show the
numeral one (1) in the left position indicating
that the cord circuit is open.