15
2.
Move the Mode selector switch to all three
positions and test the LED indication.
3.
Check the continuity between receptacle
ground and case ground using the test cable.
If necessary adjust the display to zero. When
the open ground switch is pressed down the
connection will be open.
4.
Connect the Test Lead to the TEST JACK.
Move the selector switch to CORD
RESISTANCE. The reading should be 1010
mil 5%.
5.
Move the switch to CASE LEAKAGE. The
display will indicate 200 Micro Amps + 2%.
These tests will confirm that the LKG 601 is
working properly.