Chapter 7: Product Specification
Electrical
AC Power:
90-264 VAC, 50-60 Hz, 1.0A/115V
–
0.5A/230V
Type of Protection against
Electrical Shock:
Class II Equipment
Degree of Protection against
Electrical Shock:
Type B.F. Applied Part
Degree of Protection against
Ingress of Water:
IP23
Mode of Operation:
Continuous
Physical
Device Dimensions
23.94” L x 12.25” W x 1.91” H (60.81 x 31.12 x 4.85 cm)
Device Weight
< 11 lbs. (5 Kg)
Patient Area
21.41” L x 10.24” W (54.38 x 26 cm)
Effective Surface Area
119in2 (769 cm2 )
Light Distribution Ratio (min to
max)
>0.4 (avg. 0.79)
Controller Dimensions
4.75” L x 2.75” W x 0.89” H (12.07 x 7 x 2.26 cm)
Controller Weight
<0.5 lb. (0.23 Kg)
Length of device cord to
Controller
20" (50.8 cm)
Length of connecting cord
between controller and power
adapter
53" (134.62 cm)
Length of the power cord from
the adapter to plug
66.5" (168.91 cm)
Transportation and Storage
The Skylife device should be transported and stored in temperatures between -13°F and 122°F (-25°C
and 50°C), atmospheric pressures between 50 and 106 kPa (7.25 psi to 15.37 psi) and relative humidity
between 10% and 93% R.H.
Operating Conditions
The Skylife system should be operated in temperatures between 50°F and 98.6°F (10°C and 37°C),
atmospheric pressures between 50 and 106 kPa (7.25psi to 15.3psi), and relative humidity between
15% and 90% R.H.
Stabilization Time
The Skylife Phototherapy System does not require a pre-aging time. It is ready for use immediately after
turning on the initial time. The Skylife Phototherapy System does not require a stabilization period. It is
ready for use immediately after it is powered ON
Calibration
No calibration is required to be performed by the user.