![NeoLight SKYLIFE Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/neolight/skylife/skylife_service-manual_719081034.webp)
© 2022 by NeoLight LLC.
All rights reserved.
Page 34 of 43
500050 Rev A
System Indicators
Likely Cause
Trouble Shooting Steps
Functionality Issues
- Warning Message:
"!! SYSTEM FAILURE
00064"
Bank 2 EEPROM error
1.
Restart the system
2.
If the error persists, contact NeoLight
- Warning Message:
"!! SYSTEM FAILURE
00120"
Loose Controller J1
Connector
1.
Contact NeoLight
- Warning Message:
"!! SYSTEM FAILURE
00128"
Stalled Fan
1.
Restart the system
2.
If the error persists, contact NeoLight
- Warning Message:
"!! SYSTEM FAILURE
00256"
Temperature sensor
error
1.
Turn off the device
2.
Check if vents are blocked.
3.
Clear the area around the device and restart the device
after 10 mins.
4.
If the issue persists, contact NeoLight.
Any other error not
specified above
1.
Contact NeoLight