Page 24 of 43
500050 Rev A
𝐼
𝑚𝑒𝑎𝑛_𝑎𝑑𝑢𝑠𝑡𝑒𝑑
= 𝐼
𝑚𝑒𝑎𝑛
+
δ
7.
Compare the adjusted mean values against the specification.
Treatment Setting
Performance Specification
Acceptance Criteria
Low Setting
25.0 < I
mean_adjusted_low
≤
35.0
Pass
High Setting
35.0 < I
mean_adjusted_high
≤
55.0
Pass
Very High Setting
55.0 < I
mean_adjusted_vhigh
≤
72.4
Pass
8.
Spectral measurements are prone to variation in measurement techniques. If the device fails to meet
the requirements a 2
nd
time, contact NeoLight customer support.
4.4 Replacing the GelMat
CAUTION
:
Do not clean the Light Bed top surface with any liquids. Instead, use dry cloth wipes only
on the cloudy area of the Light Bed.
Tools and Materials Required
1.
70% Iso Propyl Alcohol
2.
Microfiber cloth
NOTE
GelMat should only be replaced by a Biomed technician or other medical professional.
Please document this replacement in the Service Guide provided in Appendices A & B.
1.
Remove the old GelMat and dispose of it per institutional practices. The GelMat is made from
Polyurethane and
cannot
be recycled.
2.
Clean the Light Bed
plastic surface with 70%
Isopropyl Alcohol and a
microfiber towel to
remove any remaining
adhesive residue. Ensure
that there is no leftover
adhesive or sticky residue
before continuing