Troubleshooting
N-550
49
Data Port
Table 10 lists a symptom for data port problems and recommended actions. If
the action requires replacement of the User Interface PCB, refer to
Error Codes
When the N
-
550 detects a defective sensor connected to the N-550, the N-550
displays an error code of “SEN Err.” The sensor should be replaced and the
N-550 power should be cycled.
An error code is displayed when the N-550 detects a non-correctable failure.
Cycling the power clears these errors. Table 11 lists the N-550 error codes and
their meanings. When one of these errors occur, the N-550:
•
sounds a low priority alarm that cannot be silenced except by
power-down
•
stops measurements
Table 10: Data Port Problems
Symptom
Recommended Action
No printout is being
received.
The N-550 is running on battery power. Connect the
N-550 to AC power.
The N-550's baud rate does not match the printer.
Change the baud rate of the N-550 following
instructions in the Operator’s manual.
Check the connection between the data port and the
printer. Refer to
Replace the Front Panel PCB. Refer to
Содержание OxiMax N-550
Страница 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Страница 10: ...Blank Page ...
Страница 38: ...Blank Page ...
Страница 64: ...Blank Page ...
Страница 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Страница 82: ...Blank Page ...
Страница 86: ...Spare Parts 76 Figure 40 Exploded View ...
Страница 90: ...Blank Page ...
Страница 108: ...Blank Page ...
Страница 120: ...Blank Page ...
Страница 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Страница 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Страница 136: ...Blank Page ...
Страница 137: ... ...