Performance Verification
16
27. Press the Pulse Rate Alarm Limit button two times rapidly. Verify that the
pulse rate display indicates an alarm limit of “40.”
28. Press the
SatSeconds
Alarm Limit button. Verify that the pulse rate
display indicates an alarm limit of “OFF.”
29. Press the Power On/Off button to turn the N-550 off.
General Operation
The following tests are check overall system performance:
•
•
LED Excitation Test
This procedure uses normal system components to test circuit operation. A
Nellcor
O
XI
M
AX
oxygen transducer (sensor), model MAX-A, is used to
examine LED intensity control. The red LED is used to verify intensity
modulation caused by the LED intensity control circuit.
1. Sensor Port
1. Connect the N-550 to an AC power source.
2. Connect a DOC-10 pulse oximetry cable to the N-550 sensor port (1).
3. Connect a MAX-A sensor to the sensor-input cable.
Содержание OxiMax N-550
Страница 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Страница 10: ...Blank Page ...
Страница 38: ...Blank Page ...
Страница 64: ...Blank Page ...
Страница 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Страница 82: ...Blank Page ...
Страница 86: ...Spare Parts 76 Figure 40 Exploded View ...
Страница 90: ...Blank Page ...
Страница 108: ...Blank Page ...
Страница 120: ...Blank Page ...
Страница 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Страница 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Страница 136: ...Blank Page ...
Страница 137: ... ...