English-14
How to Change Confi guration (Preset 1, 2, 3)
The NEC model MD215MG is shipped with 3 factory preset confi gurations, as listed below.
Confi guration
Luminance (cd/m
2
)
Gamma
Remark
Maximum
Minimum
Ambient
1
400
1.0
0.0
DICOM
[Factory Default]
2
430
1.5
0.0
DICOM
3
500
1.0
0.0
DICOM
The confi guration can be changed via an OSD menu. These parameters may be modifi ed anytime by a new display calibration.
Regarding display calibration, please refer to the
MD215MG Calibrator Manual
.
1. Click the (+) plus button or (-) minus button on the back side of the monitor twice to display the “PRESENT STATE” OSD.
A display like this picture opens on the screen.
2. Hold down the (+) plus button and the (-) minus button simultaneously for over 2 seconds.
The “SWITCHING CONFIGURATION” menu opens on the screen.
3. Select your desired confi guration by the (+) plus button or the (-) minus button.
4. After selecting your preferred CONFIGURATION mode, hold down the (+) plus button and the (-) minus button
simultaneously for over 2 seconds again. Then confi rm any change of confi guration in the “PRESENT STATE” OSD.
5. Click the (+) plus button or (-) minus button once to close the “PRESENT STATE” OSD.