background image

4

 

2.0  Patient Consultation

 

Communicating with your patient and establishing an environment where they feel cared for and comfortable is a 
crucial step to set the tone of the treatment experience.

 

PLASMA IQ is used in the removal and destruction of skin lesions and the coagulation of tissue. 

Take a moment to talk with your patient to understand:

 

Why is the patient interested in a PLASMA IQ treatment?

What indication does the patient want to address?

What level of change is the patient looking for? Discuss the amount of treatments that would most likely support
this level of change.

Share before and after imagery to help set expectations for results.

 

WARNING:

 

NOT FOR USE IN THE GENITAL AND EYE AREAS.

 

USE CAUTION AROUND SENSITIVE SKIN AREAS SUCH AS THE LIPS. 

WARNING: 
Do not directly touch the output electrode of the product with sensitive parts such as the user

s      eye or the 

patient

s eye. In addition, treatments should not be performed within the orbital 

rim, which includes the upper and lower eyelids. 

Discuss 

contraindications with your patient and ensure that he/she doesn’t have any pre

-existing conditions that 

would make a treatment ineffective or unsafe. PLASMA IQ is Rx Only.

 

CONTRAINDICATIONS:

 

Birthmarks and Port-wine Stains

Warts

Tattoos

Blood Coagulation (Hemophilia)

Auto-immune disorders (Lupus)

Fitzpatrick Scale 4-6

Patients with an implanted pacemaker

Patients with metal or electrically conductive
implants

Cancer

Pregnancy and breastfeeding

Cardiac Disorders

Cardiovascular Diseases

Uncontrolled Diabetes

Circulatory Disorders

Active Herpes or Shingles

Skin Disorders including psoriasis, eczema,
dermatitis, Vitiligo

Scars including hypertrophic, keloid scars or
previous occurrences of keloid scars

PRECAUTIONS: 

Consult with the patient on the following

Hyper-pigmentation

Uncontrolled Epilepsy

Anemia

Medications or herbal supplements

Eye conditions or prior surgeries/treatments

Toxins, chemical peels, retinols, and/or dermal fillers
in the treatment area

Topical Anesthetic Intolerance

Recent surgical procedures

NOTE: 

Contact lens use during and after procedure is not recommended

1

 

2

 

Содержание PLASMA IQ

Страница 1: ...TRAINING MANUAL Neauvia North America Customer Service 1 866 836 3113...

Страница 2: ...NTS 1 INTRODUCTION 3 2 PATIENT CONSULTATION 4 3 PREPARING THE DEVICE 6 4 PRE TREAMENT CARE 7 5 TECHNIQUE 8 5 1 Zig zag pattern technique 8 6 SKIN LESION 10 7 POST TREATMENT RECOMMENDATIONS 11 8 DEVICE...

Страница 3: ...is only a means of support in the correct operation of the device The PLASMA IQ device should only be used by medically licensed and certified practitioners who have undergone training conducted by a...

Страница 4: ...the orbital rim which includes the upper and lower eyelids Discuss contraindications with your patient and ensure that he she doesn t have any pre existing conditions that would make a treatment ineff...

Страница 5: ...5 Swelling bruising Scabbing crusting 2 0 Patient Consultation Skin discoloration Scarring Redness tenderness RISKS OF PLASMA IQ TREATMENT...

Страница 6: ...for What are the stages of thetreatments What types of sensations can they expect throughout the treatment What they should expect immediately after the treatment What kind of after care they should f...

Страница 7: ...device there are two buttons used to provide two different output voltages marked with inscriptions H high and L low Figure 1 Device PLASMA IQ Preparing the device and equipment for work Before starti...

Страница 8: ...8 4 0 Pre Treatment Care Pre Treatment Recommendations Avoid sun exposure or tanning beds for 2 4 weeks before treatment Ideally begin applying SPF 50 to prep skin for 2 4 weeks prior to treatment 1...

Страница 9: ...as an initial measure before increasing voltage for more intensive fulguration It is recommended to use the LOW setting whenever possible for patient safety and comfort button LOW 650 V Controlled fu...

Страница 10: ...skin using gauze and a non alcohol antiseptic 4 Dry area completely with gauze 5 Attach single use disposable electrode to the PLASMA IQ device as instructed on page 6 6 Press L button to set voltage...

Страница 11: ...e effect 2 Carefully remove numbingcream 3 Cleanse the skin using gauze and a non alcohol antiseptic 4 Dry area completely with gauze 5 Attach single use disposable electrode to the PLASMA IQ device a...

Страница 12: ...is normal and typically only lasts for about an hour though some variation in time may occur Continue to apply moisturizer such as petroleum jelly twice a day after gentle cleansing Additional moistu...

Страница 13: ...t position as described in the manual If the electrode is mounted incorrectly it may affect the performance of the device WARNING Tighten the active electrode holder securely Improper tightening of th...

Страница 14: ...visibly wet approximately 3 5 minutes Maintenance If you experience technical difficulty with your device contact Neauvia North America at 1 866 836 3113 or email service neauvia us com PLASMA IQ dev...

Страница 15: ...14...

Страница 16: ...15...

Страница 17: ...Neauvia North America Customer Service 1 866 836 3113 2019 Neauvia North America 8480 Honeycutt Road Raleigh NC 27615 neauvia com PI NAM 01 190005 REV Code 122021...

Отзывы: