8.0 Device  Maintenance

 

13 

Cleaning and disinfection

 

Plasma IQ device does not directly contact the patient. The Operator of the Plasma IQ device is advised to use
proper protective equipment including gloves. The Plasma IQ device should be cleaned before and after each
use. The active electrode is for single use only and must be discarded after the treatment.

Do not use abrasives and free chlorine and oxygen - they may damage the case.

Clean the device with a cleaning product such as 70:30 IPA or facility approved cleaning agent for 3-5 minutes
and disinfect with a disinfecting product such as with Caviwipes or facility approved disinfectant for 3-5 minutes.
Ensure that the device remains wet per the specifications of the chosen disinfectant. Refer to the cleaning and
disinfection label for specific time frames. Allow to air dry until no longer visibly wet (approximately 3-5 minutes).

Maintenance

 

If you experience technical difficulty with your device, contact Neauvia North America at 1-866-836-3113 or email 
[email protected]. PLASMA IQ devices in the United States and Canada should not be serviced by third 
party technicians. The warranty is void if the medical device has been opened, tampered with or repaired by an 
unauthorized technician. Neauvia North America will arrange to supply the customer a means to ship the defective 
device to the Neauvia North America service department for necessary repairs and will ensure that the customer is 
supplied a temporary device for use during the repair process in a timely manner.

 

The device is not equipped with any fuses that can be replaced by third party operators or service personnel. There 
are no serviceable parts inside the device. Service Personnel must hold a valid certificate issued by manufacturer to 
perform any repairs of this equipment.

 

No parts of the equipment should be serviced or maintained while in use with a patient.

 

3

 

4

 

Содержание PLASMA IQ

Страница 1: ...TRAINING MANUAL Neauvia North America Customer Service 1 866 836 3113...

Страница 2: ...NTS 1 INTRODUCTION 3 2 PATIENT CONSULTATION 4 3 PREPARING THE DEVICE 6 4 PRE TREAMENT CARE 7 5 TECHNIQUE 8 5 1 Zig zag pattern technique 8 6 SKIN LESION 10 7 POST TREATMENT RECOMMENDATIONS 11 8 DEVICE...

Страница 3: ...is only a means of support in the correct operation of the device The PLASMA IQ device should only be used by medically licensed and certified practitioners who have undergone training conducted by a...

Страница 4: ...the orbital rim which includes the upper and lower eyelids Discuss contraindications with your patient and ensure that he she doesn t have any pre existing conditions that would make a treatment ineff...

Страница 5: ...5 Swelling bruising Scabbing crusting 2 0 Patient Consultation Skin discoloration Scarring Redness tenderness RISKS OF PLASMA IQ TREATMENT...

Страница 6: ...for What are the stages of thetreatments What types of sensations can they expect throughout the treatment What they should expect immediately after the treatment What kind of after care they should f...

Страница 7: ...device there are two buttons used to provide two different output voltages marked with inscriptions H high and L low Figure 1 Device PLASMA IQ Preparing the device and equipment for work Before starti...

Страница 8: ...8 4 0 Pre Treatment Care Pre Treatment Recommendations Avoid sun exposure or tanning beds for 2 4 weeks before treatment Ideally begin applying SPF 50 to prep skin for 2 4 weeks prior to treatment 1...

Страница 9: ...as an initial measure before increasing voltage for more intensive fulguration It is recommended to use the LOW setting whenever possible for patient safety and comfort button LOW 650 V Controlled fu...

Страница 10: ...skin using gauze and a non alcohol antiseptic 4 Dry area completely with gauze 5 Attach single use disposable electrode to the PLASMA IQ device as instructed on page 6 6 Press L button to set voltage...

Страница 11: ...e effect 2 Carefully remove numbingcream 3 Cleanse the skin using gauze and a non alcohol antiseptic 4 Dry area completely with gauze 5 Attach single use disposable electrode to the PLASMA IQ device a...

Страница 12: ...is normal and typically only lasts for about an hour though some variation in time may occur Continue to apply moisturizer such as petroleum jelly twice a day after gentle cleansing Additional moistu...

Страница 13: ...t position as described in the manual If the electrode is mounted incorrectly it may affect the performance of the device WARNING Tighten the active electrode holder securely Improper tightening of th...

Страница 14: ...visibly wet approximately 3 5 minutes Maintenance If you experience technical difficulty with your device contact Neauvia North America at 1 866 836 3113 or email service neauvia us com PLASMA IQ dev...

Страница 15: ...14...

Страница 16: ...15...

Страница 17: ...Neauvia North America Customer Service 1 866 836 3113 2019 Neauvia North America 8480 Honeycutt Road Raleigh NC 27615 neauvia com PI NAM 01 190005 REV Code 122021...

Отзывы: