9
5.1 Technique: Zig-Zag Pattern Technique
Zig-zag pattern technique protocol
1.
Apply short energy pulses in non-linear pattern working in a zig-zag motion to ensure distribution of energy
pulses for patient comfort.
2.
Pulses should be at least 1mm apart.
Zig-zag pattern technique step-by-step
1.
Apply numbing cream to treatment area as prescribed by physician and allow 10-20 min to take effect.
2.
Carefully remove numbing cream.
3.
Cleanse the skin using gauze and a non-alcohol antiseptic.
4.
Dry area completely with gauze.
5.
Attach single use, disposable electrode to the PLASMA IQ device as instructed on page 6.
6.
Press L button to set voltage parameters to low.
7.
Apply short energy pulses in non-linear pattern working in a zig-zag motion to ensure distribution of energy
pulses for patient comfort.
8.
Pulses should be at least 1mm apart.
9.
Continue until the full treatment area is covered.
10.
When area is complete, move to next area and repeat steps 2-9.
11.
Cleanse treated area with gauze and non-alcohol antiseptic.
12.
Apply moisturizer such as petroleum jelly as directed by physician.