background image

 

EW0826A4_en - rev. A 

Oxygen detectors 

 URS20SS 

05/10/2020 

Gas detection systems for industrial environments 

7/8

 

 

Checklist after 
mechanical and 
electrical installation 
(continued)

 

Since the sensor scale is between 0…...30% O

2

,  in  air (20.9% O

2

) the 4…...20mA 

output must be set between 14 and 16mA (theoretic value 15.1mA). 

Use  the  calibration  kit  to  apply  the  gas  mixture  at  4%  of  CO

2

,  making  sure  the 

4...20mA output is between 5 and 7mA (theoretic value 6.1mA); the status LED and 
the pre-alarm, 1

st

 and 2

nd

 alarm threshold relays on the optional relay card switch on 

according to the thresholds and the alarm mode settings. 
 

Maintenance

 

Routine 
 

Corrective 

 
 
 
 
 
 

Decommissioning

 

A sensor functional test should be carried out every three-six months. 

Routine maintenance involves repeating the same tests as set forth in “checklist after 
mechanical and electrical 

installation”.

 

If any abnormalities are found during routine sensor maintenance, return the sensor 
concerned to the supplier / installer, who in turn will send it back to the manufacturer. 
Sensors  may  need  to  be  re-calibrated,  using  the 

TUL40..

  calibration  kit  and  the 

TUS40-20

  handheld  terminal,  which  must  be  connected  to  the  sensor  via  the 

communication interface integrated in the cable (on the connector 

CN4

). 

 

For  the  re-calibration  procedure,  see  the  instructions  supplied  with  the  handheld 
terminal. 

Remove  power from the detector, disconnect all wiring and conduits and dismount 
the housing from all the blocking systems. 
 

Warranty 

Warranty on EsiWelma products is valid for 12 months from installation date and no 
longer  than  24  months  from  manufacturing  date  on  the  product.  Installation  data, 
stamp  and  signature  on  the  data  sheet  filled  in  by  the  installer  will  be  considered 
proof for warranty. 

 

A copy of the warranty data sheet must be sent when returning the product under 
warranty. 

 

Accessories 

UZR20.4

 Four-relay card 

TUL40..

 Gas calibration Kit 

TUS40

 Handheld terminal 

CRG40

 Gas collecting cone 

PAP40

 Powerful jets protection 

 

Dimensions and weight 

 

Dimensions (HxWxD): 155x100x60mm.                                              Weight: 0.65Kg  
 

 

 

Содержание EsiWelma Sensigas URS20SS

Страница 1: ...eaks in hospitals laboratories welding centres and more in general where oxygen is stored Oxygen deficiency is an indirect measurement of the presence of other explosive or asphyxiant gases that deple...

Страница 2: ...umption mode applications at 1 or 2 thresholds 4 20mA Output reference selection 4mA 0 O2 20mA 30 O2 0mA no alarm 10mA 1st threshold alarm 20mA 2nd threshold alarm by jumper selectable polarity 4 20mA...

Страница 3: ...nstructions for location of the detectors The detectors must be installed 1 where accidental gas leakages are possible 2 at least 1 5m away from heat sources or from vent holes 3 not in spaces where v...

Страница 4: ...n order to change the default settings switch off the power supply input the new settings at the JP2 jumper circuit or at the S1 DIP switch as shown in the diagram then power up again in particular Th...

Страница 5: ...Step 1 Insert the connection slot provided with the relay card into the control card making sure the flexible tab is towards the main terminal board Find CN3 connector CN3 connector Connection slot fl...

Страница 6: ...on The sensors are factory calibrated so they normally do not require any other calibration once installed Still after installation an operational check of the sensors is recommended The detector will...

Страница 7: ...return the sensor concerned to the supplier installer who in turn will send it back to the manufacturer Sensors may need to be re calibrated using the TUL40 calibration kit and the TUS40 20 handheld t...

Страница 8: ...gas vapours flammable mist Ex nA nC d IIC T6 Gb Types of protection complying with EN60079 0 and EN60079 15 sensor body with type of protection d in compliance with EN60079 1 BVI 07 ATEX 0033 Type exa...

Отзывы: