Natus Photic Stimulator

 

 

User & Service Manual

  

 

16 

IEC 60601-1-2:2007 Table 4 Requirements: 

 

The Natus Photic Stimulator is intended for use in the electromagnetic environment specified 

below. The customer or the user of the equipment should assure that it is used in such an 

environment. 

Immunity test 

IEC 60601 test 

level 

Compliance 

level 

Electromagnetic environment 

– guidance 

 
 

Conducted RF 
IEC 61000-4-6  

 
 
 
 
 
 
 
 
 

Radiated RF IEC 

61000-4-3 

 
 

3 Vrms  

150 kHz to 80 

MHz  

 
 
 
 
 
 
 
 
 

3 V/m 80 MHz 

to 2,5 GHz 

 
 
 

3 Vrms  

 
 
 
 
 
 
 
 
 

1 V/m 

 

Portable and mobile RF communications equipment 
should be used no closer to any part of the equipment 
including cables, than the recommended separation 
distance calculated from the equation applicable to 
the frequency of the transmitter.  

 

Recommended separation distance 

 

where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation 
distance in meters (m). Field strengths from fixed RF 
transmitters, as determined by an electromagnetic 
site survey

a

 should be less than the compliance level 

in each frequency range

b

 

 
 Interference may occur in the vicinity of known RF 
transmitting devices and equipment marked with the 
following symbol: 

 

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.  
 
NOTE  2  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by  absorption  and 
reflection from structures, objects and people. 

a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the equipment is used exceeds the applicable RF 
compliance level above, the equipment should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as re-orienting or relocating the equipment 
 
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

 

 

Содержание Photic Stimulator

Страница 1: ...Natus Photic Stimulator User Service Manual...

Страница 2: ......

Страница 3: ...use extend to all media in which this information is preserved This copy of the User Manual shall be used only in accordance with the conditions of sale of Natus Neurology Incorporated or its distribu...

Страница 4: ...ENDED USER PERFORMED MAINTENANCE 10 Cleaning 10 Inspection 10 Maintenance 10 ENVIRONMENTAL SPECIFICATIONS 11 Temperature 11 Humidity 11 Condensation 11 Flammability 11 EMC Specifications 11 Safety Sta...

Страница 5: ...by the Natus NeuroWorks TM software from laptop or desktop computers it supplies intense flashes of light Its components consist of an arm mounted light assembly a light source an intensity control a...

Страница 6: ...presence of flammable anesthetics Reliable grounding requires hospital grade receptacles and power cord Always perform a leakage current test and compare to allowable standards BEFORE connecting the p...

Страница 7: ...test In addition the accompanying EEG Electroencephalograph amplifier s signals will also be contaminated past the point where any clinical signal interpretation is possible Trained electroencephalog...

Страница 8: ...on of warnings and markings in a restricted space ATTENTION Consult Accompanying Documents Consult Accompanying Documents Protective Earth Ground Type B equipment Type BF Equipment Dangerous Voltage A...

Страница 9: ...ort arm 52 in 1320 8 mm total Frequency of flash Maximum 60 Hz Light intensity Adjustable via Intensity control 12 position control 11 settings 1 Off Input requirements TTL Positive Pulse 100 s 1 mA M...

Страница 10: ...Stimulator Netlink Cable 580 PHTCB3 Photic Stimulator XL Cable Installation and Operation WARNING Never place the photic device on the floor 1 Place the photic device on a platform table cart or other...

Страница 11: ...onnector 8 On the Gamma headbox connect the phone plug of trigger cable BLSC part number 580 PHTCB2 labeled Strobe Out to the phone jack on back panel of computer labeled Strobe 9 Use the photic devic...

Страница 12: ...pecial cleaning is required for this device Ordinary house keeping type cleaning to prevent the accumulation of dust or debris is adequate Inspection No special inspection procedures are required for...

Страница 13: ...ability UL94 HB UL94V O EMC Specifications IEC 60601 1 2 2007 Electromagnetic Compatibility EMC Testing CISPR 11 B Conducted and CISPR 11A Radiated Emissions EN 61000 4 2 Electrostatic Discharge Test...

Страница 14: ...anual Room Topography Place all equipment on an even level surface Avoid the potential for mechanical shock or possible vibrations during setup system operation or when relocating the equipment Verify...

Страница 15: ...ients are defined as short bursts of energy that are propagated through the power cord The EFT source is usually located in the nearby equipment or machinery such as motors switching equipment light s...

Страница 16: ...rence in nearby electronic equipment RF emissions CISPR 11 Class A The Natus Photic Stimulator is suitable for use in all establishments other than domestic and may be used in domestic establishments...

Страница 17: ...ment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltag...

Страница 18: ...electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of known RF transmitting devices and equipment marked with the fol...

Страница 19: ...mulator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Natus Photic Stimulator as recommen...

Страница 20: ...Natus Photic Stimulator User Service Manual 18 Intentionally Left Blank...

Страница 21: ...User Service Manual Natus Photic Stimulator 19 Intentionally Left Blank...

Страница 22: ...Natus Photic Stimulator User Service Manual 20 Intentionally Left Blank...

Страница 23: ...Intentionally Left Blank...

Страница 24: ...t is Natus Medical Incorporated an internationally respected innovator of medical products and services Our Neurology systems are backed up by an in house support team staffed with technical and clini...

Отзывы: