Natus Photic Stimulator

 

 

User & Service Manual

  

 

12 

Safety Standards 

This component complies with the following electrical safety standards:  

UL 60601-1 

CAN/CSA C22.2 No. 601-1-M90 

IEC 60601-1-2:2007 Electromagnetic Compatibility (EMC) Testing 

 

Environmental Conditions 

 

Select a room with properly grounded power sources. 

 

 

Do not use or store your equipment in places where chemicals are stored or where there is a 
potential for gas leakage. 

 

 

Avoid moisture or contact with water, extreme atmospheric pressure, excessive humidity and 
temperature, poorly ventilated areas and dusty, saline or sulfuric air. 

 

 

Verify the selected site maintains a relative humidity between 30% and 75% (without 
condensation). 

 

 

Verify all conditions meet the requirements listed in the Environmental Specifications section 
of this manual. 

 

Room Topography 

 

Place all equipment on an even, level surface. Avoid the potential for mechanical shock or 
possible vibrations during setup system operation or when relocating the equipment. 

 

 

Verify the power supply and all portable multiple socket-outlets are off the floor and in a  
dry location.

 

Electrostatic Discharge (ESD) Handling Procedures  
and Warnings 

Before performing any setup or placement procedures, it is recommended that all clinical staff 
read and/or are trained on the precautions outlined in this section.  
 
 

 

WARNING:

 Be sure to take the appropriate Electrostatic 

Discharge (ESD) precautions. Disconnect the cables before 
moving, cabling, or performing any set up procedures. 

 
Some semiconductor (solid state) devices can be easily damaged by static electricity. Such 
components are commonly called Electrostatically Sensitive Devices (ESD). Do not touch the 
accessible conductive parts or pins for the Connectors marked with the ESD symbol unless ESD 
precautionary procedures are used. 

Содержание Photic Stimulator

Страница 1: ...Natus Photic Stimulator User Service Manual...

Страница 2: ......

Страница 3: ...use extend to all media in which this information is preserved This copy of the User Manual shall be used only in accordance with the conditions of sale of Natus Neurology Incorporated or its distribu...

Страница 4: ...ENDED USER PERFORMED MAINTENANCE 10 Cleaning 10 Inspection 10 Maintenance 10 ENVIRONMENTAL SPECIFICATIONS 11 Temperature 11 Humidity 11 Condensation 11 Flammability 11 EMC Specifications 11 Safety Sta...

Страница 5: ...by the Natus NeuroWorks TM software from laptop or desktop computers it supplies intense flashes of light Its components consist of an arm mounted light assembly a light source an intensity control a...

Страница 6: ...presence of flammable anesthetics Reliable grounding requires hospital grade receptacles and power cord Always perform a leakage current test and compare to allowable standards BEFORE connecting the p...

Страница 7: ...test In addition the accompanying EEG Electroencephalograph amplifier s signals will also be contaminated past the point where any clinical signal interpretation is possible Trained electroencephalog...

Страница 8: ...on of warnings and markings in a restricted space ATTENTION Consult Accompanying Documents Consult Accompanying Documents Protective Earth Ground Type B equipment Type BF Equipment Dangerous Voltage A...

Страница 9: ...ort arm 52 in 1320 8 mm total Frequency of flash Maximum 60 Hz Light intensity Adjustable via Intensity control 12 position control 11 settings 1 Off Input requirements TTL Positive Pulse 100 s 1 mA M...

Страница 10: ...Stimulator Netlink Cable 580 PHTCB3 Photic Stimulator XL Cable Installation and Operation WARNING Never place the photic device on the floor 1 Place the photic device on a platform table cart or other...

Страница 11: ...onnector 8 On the Gamma headbox connect the phone plug of trigger cable BLSC part number 580 PHTCB2 labeled Strobe Out to the phone jack on back panel of computer labeled Strobe 9 Use the photic devic...

Страница 12: ...pecial cleaning is required for this device Ordinary house keeping type cleaning to prevent the accumulation of dust or debris is adequate Inspection No special inspection procedures are required for...

Страница 13: ...ability UL94 HB UL94V O EMC Specifications IEC 60601 1 2 2007 Electromagnetic Compatibility EMC Testing CISPR 11 B Conducted and CISPR 11A Radiated Emissions EN 61000 4 2 Electrostatic Discharge Test...

Страница 14: ...anual Room Topography Place all equipment on an even level surface Avoid the potential for mechanical shock or possible vibrations during setup system operation or when relocating the equipment Verify...

Страница 15: ...ients are defined as short bursts of energy that are propagated through the power cord The EFT source is usually located in the nearby equipment or machinery such as motors switching equipment light s...

Страница 16: ...rence in nearby electronic equipment RF emissions CISPR 11 Class A The Natus Photic Stimulator is suitable for use in all establishments other than domestic and may be used in domestic establishments...

Страница 17: ...ment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltag...

Страница 18: ...electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of known RF transmitting devices and equipment marked with the fol...

Страница 19: ...mulator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Natus Photic Stimulator as recommen...

Страница 20: ...Natus Photic Stimulator User Service Manual 18 Intentionally Left Blank...

Страница 21: ...User Service Manual Natus Photic Stimulator 19 Intentionally Left Blank...

Страница 22: ...Natus Photic Stimulator User Service Manual 20 Intentionally Left Blank...

Страница 23: ...Intentionally Left Blank...

Страница 24: ...t is Natus Medical Incorporated an internationally respected innovator of medical products and services Our Neurology systems are backed up by an in house support team staffed with technical and clini...

Отзывы: