C H A P T E R 2
R
OUTINE
M
AINTENANCE
P
ROCEDURES
This chapter includes procedures for routine maintenance of the neoBLUE mini device,
hereafter called the UUT (Unit Under Test):
2.1 Cleaning
2.2 Safety
Testing
2.3 Performance
Verification
These procedures must be performed after any of the following events and/or as prescribed by
the institution’s maintenance schedule.
Initial receipt of the UUT at the institution.
Foreign material or liquid has been deposited on the UUT.
UUT has been physically damaged or subjected to mechanical shock (i.e., being
dropped).
UUT has been submitted for maintenance or scheduled performance verification.
2.1 Cleaning
Cleaning the UUT incorporates two different tasks:
Cleaning the Light Enclosure and Adjustable Arm (Section 2.1.1)
Cleaning the Interior of the Light Enclosure (Section 2.1.2)
Warning! Disconnect light from AC power before cleaning.
2.1.1 Cleaning the Light Enclosure and Adjustable Arm
Remove dust from the exterior of the device with a soft brush or soft cloth dampened with water.
Sponge away remaining debris with a mild solution of detergent and water, a non-caustic
commercial cleaner, or hospital disinfectant.
Clean the lens with a soft cloth dampened with water. If water alone is ineffective in removing
fingerprints or other markings, use a mild solution of detergent and water, a non-caustic
commercial cleaner, or hospital disinfectant.
Caution: Observe the following precautions:
Do not spray liquids directly onto the light, or allow them to seep into the interior.
Do not use caustic or abrasive cleaners.
Do not clean with alcohol, acetone, or other solvents.
Never immerse the light or its component parts.
neoBLUE
®
mini System Service Manual
P/N 051466F
3